BELIMED STEAM STERILIZER MST-H

Sauter AG

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEJune 17, 2015PRODUCT CODEFLE

Indications for use

The Belimed Steam Sterilizers MST-H, models 9- 6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable material s used in healthcare facilities. The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version. The Belimed Steam Sterilizer MST-H is offered with the following factory-programmed predefined sterilization cycles: Cycle Pre-treatment Sterilizing temperature Sterilizing Time 1) Drying time 2) Recommended load 3) PreVac 270 4S/30D 5 pulses 270 °F 4 min 30 min Double wrapped instrument trays, max. weight of 25 lbs per tray, Fabric packs PreVac 270 4S/5D 5 pulses 270 °F 4 min 5 min Fabric packs 4) PreVac 275 3S/30D 5 pulses 275 °F 3 min 30 min Double wrapped instrument trays, max. weight of 25 lbs per tray, Fabric packs Bowie-Dick Test 5 pulses 273 °F 3.5 min 1 min One Bowie-Dick-Test-Pack Gravity 270 15S/30D Purge 270 °F 15 min 30 min Double wrapped instrument trays, with non-porous instruments, max. weight of 25 lbs per tray Liquid 250 45S Purge 250 °F 45 min 0 min Liquids, up to 1000 ml in vented or open containers 5) Air leak Test Vacuum: 65mbar Test time: 15 min

- - - Empty chamber Warm up & Air Leak Test Vacuum: 65mbar Test time: 15 min 270 °F 3 min 3 min Empty chamber Warm up 2 pulses 270°F 3 min 3 min Empty chamber Table 1: Factory programmed Sterilization cycles Notes for Table 1:

1. Only factory set sterilizing time should be utilized with medical devices intended for patient contact.

2. Factory set drying time. These are the minimum validated drying times, which may be extended depending on load conditions and wraps.

3. Recommended load: Refer to table 2 and 3. The load configurations listed in table 2 and 3 are those used during performance testing of the sterilizer models.

4. Fabric load should be preconditioned between 68°F and 75°F and at a relative humidity of at least 35% to 60% for at least 2 hours.

5. Liquid cycle is not intended to sterilize liquids that are used for direct patient contact. The following tables show the recommended loads for the Belimed Steam Sterilizer MST-H: Model Instrument trays, max. 25 lbs each Fabric Packs, max. 3.3 lbs each 9-6-12 16 32 9-6-15 20 40 9-6-18 24 48 Table 2: Recommended loads for non-porous and porous load cycles Model Volume of liquid in one container Number of containers 9-6-12 1000 ml 168 9-6-15 1000 ml 210 9-6-18 1000 ml 252 Table 3: Recommended loads for liquid cycle 250 °F The Belimed Steam Sterilizer MST-H is offered in the following models: Model Configuration Net loading capacity (L) Chamber size (H x W x D) (mm) Overall Dimensions (H x W x D) (mm) 9-6-12 HS1 1 door 691 1080 x 660 x 1398 1970 x 1700 x 1714 9-6-12 HS2 2 doors 691 1080 x 660 x 1398 1970 x 1700 x 1714 9-6-15 HS1 1 door 864 1080 x 660 x 1706 1970 x 1700 x 2022 9-6-15 HS2 2 doors 864 1080 x 660 x 1706 1970 x 1700 x 2022 9-6-18 HS1 1 door 977 1080 x 660 x 2014 1970 x 1700 x 2330 9-6-18 HS2 2 doors 977 1080 x 660 x 2014 1970 x 1700 x 2330 Table 4: Models and dimensions

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。