FDA DEVICE FACTS
Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs
Maxter Glove Manufacturing Sdn Bhd
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner.
In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Test Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time (Minutes) Cisplatin, 1.0mg/ml >240 Cyclophosphamide (Cytoxan), 20mg/ml >240 Carmustine (BCNU), 3.3mg/ml 1.0 Dacarbazine (DTIC), 10.0mg/ml >240 Doxorubicin Hydrochloride, 2.0mg/ml >240 Etoposide (Toposar), 20.0mg/ml >240 Fluorouracil, 50.0mg/ml >240 Paclitaxel (Taxol), 6.0mg/ml >240 Thiotepa, 10.0mg/ml 31.3 Methotrexate ,25mg/ml >240 Mitomycin C, 0.5mg/ml >240 Vincristine Sulfate, 1.0mg/ml >240 Ifosfamide, 50.0mg/ml >240
Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml: 1.0 minute Thiotepa, 10.0mg/ml: 31.3 minutes
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。