RARA CRAFT DATABASE for FDA Medical Devices
510(k)K152241

FDA DEVICE FACTS

Medtronic Transportation/Sterilization Cassettes

Medtronic Sofamor Danek USA, Inc.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEJanuary 20, 2016PRODUCT CODEKCT

Indications for use

The Medtronic Transportation/Sterilization Cassettes are intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.

Sterilization validations for the worst case Medtronic Transportation/Sterilization Cassette (22.75 x 11.26 x 5.5 inches) included implants and common surgical instruments such as rasps, drivers, trials, handles, inserters, probes, drills, etc. The validated total weight was 28.4lbs. The validated worst case loading configurations of the Medtronic Transportation/ Sterilization Cassette included the following worst case lumen dimensions.

- 363 x 1.575mm

- 247.5 x 4.1mm

Device List: 1850060 Case - Triple Generic Outer Base (22.74 x 11.260 x 5.040 inches) 1850064 Lid - Generic Outer Lid (22.75 x 11.260 x 0.470 inches) 7022101L Tray Lid (21 x 10.13 x 0.075 inches) P1850061 Tray 1 (20.75 x 9.79 x 1.32 inches) P1850062 Tray 2 (21 x 10.13 x 1.69 inches) P1850063 Tray 3 (21 x 10.13 x 1.38 inches) 7059532 Large Caddy (9.47 x 6.37 x 1.3 inches) 7059532L Large Lid (5.85 x 4.725 x 0.095 inches) P9213018 Small Caddy (2 x 1.5 x 1.025 inches) P9213018 Small Lid (2 x 1.29 x 0.095 inches)

Sterilization Parameters: Cycle Temperature Exposure time Minimum dry time Gravity Displacement 250°F (121°C) 30 Minutes 30 Minutes Gravity Displacement 270°F (132°C) 15 Minutes 30 Minutes Gravity Displacement 275°F (135°C) 10 Minutes 30 Minutes Dynamic-Air-Removal (4 Pre-conditioning pulses) 270°F (132°C) 4 Minutes 30 Minutes Dynamic-Air-Removal (4 Pre-conditioning pulses) 275°F (135°C) 3 Minutes 30 Minutes

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン