CS-Medical TD 100 Disinfector with TD-5 and TD-8 High Level Disinfectants, CS Medical TD-5 High Level Disinfectant, CS Medical TD-8 High Level Disinfectant

Cs Medical, LLC

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATENovember 3, 2016PRODUCT CODEPSW

Indications for use

The TD 100® disinfector is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-5® or TD-8® disinfectants, which are designed to be used only with the TD 100® disinfector. The disinfectant bottles cannot be reused in the system.

TD-5® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 100® disinfector for high-level disinfection of TEE ultrasound probes.

TD-5® high level disinfectant and TD 100® disinfector system are intended for use by qualified individuals trained in its use.

TD-5® disinfectant should be used with the following contact conditions in TD 100® disinfector:

High-Level Disinfectant Time Temperature Minimum Recommended Concentration TD-5® 5 minutes 38°C — 40°C 1.7% glutaraldehyde

TD-8® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 100® disinfector for high-level disinfection of TEE ultrasound probes.

TD-8® high level disinfectant and TD 100® disinfector system arc intended for use by qualified individuals trained in its usc.

TD-8® disinfectant should be used with the following contact conditions in TD 100® disinfector:

High-Level Disinfectant Time Temperature Minimum Recommended Concentration TD-8® 5 minutes 38°C — 40°C 0.3% ortho-phthalaldehyde

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: September 2, 2026