FDA DEVICE FACTS
Matrix Pre-Treatment Controller
Better Water, LLC
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
Matrix Pre-Treatment Controller is designed to safely and effectively provide monitoring and control of the pre-treatment components for water purification for dialysis applications.
These pre-treatment components include media-tanks with controllable valves (for example, but not limited to: multi-media depth filters, filox media filters, water softeners, dealkalizers, organic scavengers, and carbon media filters), city booster pump, ultra-violet light dump valve, blend-valve by-pass valve, and the main water shut-off valve.
It monitors water pressures, water temperature, Delta-P pressures (change in pressure between two points), water flow-rate, water pH, empty bed contact time for carbon media filters, and the sta tus of the media-tanks' valve cycles. This operational data is displayed on a HMI-touch screen along with trends, alarms, and events.
It controls the regeneration and backwash cycles for media tanks, the city booster pump, the ultra -violet light dump valve for cooling purposes, the blend-valve by-pass valve to allow greater flow during backwash and regeneration, and the main water shut-off valve if a leak is detected.
Although it is designed to work with Better Water LLC equipment, it can be used with no n-Better Water equipment provided certain components are added or replaced to ensure compatibility.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。