Steri-Cassette and Steri-Cage Sterilization Packaging System

Sybron Dental Specialties

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEApril 21, 2017PRODUCT CODEKCT

Indications for use

The Steri-Cassettes and Steri-Cages are intended to contain dental instruments for cleaning, sterilization, organization, storage and handling. The Steri-Cassettes and Steri-Cages are suitable for pre-vacuum and gravity displacement steam sterilization methods. The cassettes and cages are not intended to maintain sterility; they are intended to be used in conjunction with validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices.

The validated steam sterilization cycle parameters are as follows:

Steam Sterilization Cycle Minimum Exposure Temperature Minimum Exposure Time Minimum Dry Time Gravity Displacement (wrapped) 250°F (121°C) 30 Minutes 15 Minutes Gravity Displacement (wrapped) 270°F (132°C) 15 Minutes 15 Minutes Gravity Displacement (wrapped) 275°F (135°C) 10 Minutes 30 Minutes Gravity Displacement (unwrapped for immediate use) 132°C (270°F) 3 Minutes None

Pre-Vacuum (wrapped) 270°F (132°C) 4 Minutes 30 Minutes Pre-Vacuum (wrapped) 275°F (135°C) 3 Minutes 16 Minutes Pre-Vacuum PolyVac's delivery systems consist of perforated trays with lids, which are intended to enclose and protect medical device instrumentation, and to facilitate the sterilization process by allowing steam penetration and air removal. When used in conjunction with an approved sterilization wrap, sterility of the enclosed medical device is maintained until used. PolyVac's delivery systems are to be sterilized in one of the following cycles: Prevacuum Steam: 132°C - 4 minutes minimum Gravity Steam: 132°C - 30 minutes minimum Gravity Steam: 121°C - 55 minutes minimum Similar Element Subject Device Steri-Cassette and Steri-Cage Predicate Device (K012105) PolyVac Surgical Instrument Delivery System’s Instrument Cassettes Subject Device vs.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 29, 2026