RARA CRAFT DATABASE for FDA Medical Devices
510(k)K163096

FDA DEVICE FACTS

GPC BRAND Posterior, Non cervical Pedicle Screw Spinal System

Gpc Medical Limited

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEMarch 31, 2017PRODUCT CODENKB

Indications for use

Simila r intended use in New Device and Pre dicate device Equivalent 02 Material Same m aterial used in New Device and Predic ate device Equivalent 03 Performance Standar ds Same performance standards used in both New Device as well as predicate device Equiva lent 04 Sterilization Same m ethod of sterilization used in both Ne w Device as well as Predicate device Equivalent 05 Dimensional Verific ation Same dimensions found in both New Device as well as Predicate device. Equivalent b1).Discussion on the non-clinical testing performed Following are the applicable product standards considered for non-clinical standards A:Material Standards B:Performance Standards A:Material Standards: The material standards are the essential part to be complied to first, as it is the basis of manufacturing metallic surgical implants. We have complied to following material standards

1. ASTMF136:Standard specification for wrought Titanium-6Aluminium-4Vanadium ELI (Extra low interstitial) Alloy for surgical implant applications. GPC MEDICAL LTD B:Performance Standards: The device performance of GPC BRAND Pedicle Screw System has been demonstrated against following applicable standards

1. ASTM F 1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model We have verified the compliance to these standards and copies of the relevant test results were provided in the submission. b2). Discussion on the clinical evaluation referenced and relied upon: GPC BRAND Posterior, Non cervical Pedicle Screw Spinal System are of similar design and pattern as well as similar intended use. Clinical information was not necessary to demonstrate substantial equivalence.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。