NuVasive LessRay with Enhanced Tracking

Nu Vasive, Incorporated

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEAugust 18, 2017PRODUCT CODEOWB

Indications for use

LessRay® with Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.

NuVasive® LessRay® with Enhanced Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease. Same Device Class II II Same Product Code OWB, JAA, LLZ OWB, JAA, LLZ Same Regulation Number (21CFR) §892.1650 §892.1650 Same Device Classification Name Interventional Fluoroscopic X-Ray System Interventional Fluoroscopic X-Ray System Same NuVasive® LessRay® with Enhanced Tracking Specification / Property Predicate Device Subject Device Discuss ion LessRay with Tracking (K142243)

NuVasive LessRay with Enhanced Tracking Device Functionaliti es (image acquisition, enhancement, and display)

- Software based device used to provide computer display systems interfaced to a fluoroscope through a video cable. The images produced by the fluoroscope are transmitted through a cable to a frame capture board in the computer where the images are enhanced and then displayed on the monitor.

- Enhanced images are displayed on a computer monitor at the same time that the corresponding original image is displayed on the fluoroscope monitor(s).

- Serves only as an image display which is in addition to the fluoroscope’s standard image display device. Device is passive, in that the operation depends only on the video output of the fluoroscope, and it does not transmit any signals or images to the fluoroscope.

- Software based device used to provide computer display systems interfaced to fluoroscope through a video cable. The images produced by the fluoroscope are transmitted through a cable to a frame capture board in the computer where the images are enhanced and then displayed on the monitor.

- Enhanced images are displayed on a computer monitor at the same time that the corresponding original image is displayed on the fluoroscope monitor(s).

- Serves only as an image display which is in addition to the fluoroscope’s standard image display device. Device is passive, in that the operation depends only on the video output of the fluoroscope, and it does not transmit any signals or images to the fluoroscope. Same Algorithms

- Image quality improvement using averaging algorithm

- Contrast and brightness enhancement with simultaneous reduction of random noise

- Image quality improvement using averaging algorithm

- Contrast and brightness enhancement with simultaneous reduction of random noise Same Compatible hardware platforms Any computer that meets the following minimum specifications: CPU: Intel Core 2 Duo GPU: NVIDIA Quadro 4000 RAM: 8 GB HDD: 256 GB Frame Grabber: Aver Media H339 or Elgato Operating System: Windows 7 or 8.1 Any computer that meets the following minimum specifications: CPU: Intel Core 2 Duo GPU: NVIDIA Quadro 4000 RAM: 8 GB HDD: 256 GB Frame Grabber: Aver Media H339 or Elgato Operating System: Windows 8.1 same NuVasive® LessRay® with Enhanced Tracking Specification / Property Predicate Device Subject Device Discuss ion LessRay with Tracking (K142243)

NuVasive LessRay with Enhanced Tracking C-arm Tracking

- When tracking is enabled, will automatically choose the Baseline when the fluoroscope is near the location and orientation that the Baseline was initially taken.

- When tracking is enabled, requires hardware components in order to mount the off-the-shelf tracking hardware to the C-arm and to the operating table.

- When tracking is enabled, requires the use of an off-the-shelf tracking system in order to track the 6 DOF location of the C-arm relative to the operating table.

- When tracking is enabled, visual cues are provided which help guide the user in positioning the C-arm back to where a prior Baseline was taken.

- When tracking is enabled, will automatically choose the Baseline when the fluoroscope is near the location and orientation that the Baseline was initially taken.

- When tracking is enabled, requires hardware components in order to mount the off-the-shelf tracking hardware to the C-arm and to the operating table.

- When tracking is enabled, requires the use of an off-the-shelf tracking system in order to track the 6 DOF location of the C-arm relative to the operating table.

- When tracking is enabled, visual cues are provided which help guide the user in positioning the C-arm back to where a prior Baseline was taken. Same Tracking options Electromagnetic or optical Electromagnetic or optical Same Instrument Tracking No Yes- NuVasive LessRay with Enhanced Tracking has additional capability of instrument tracking to aid the user in positioning an instrument using prior baseline x-rays. Additio nal function ality

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 29, 2026