FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEJuly 11, 2017PRODUCT CODEIYN

Indications for use

The SuperSonic Imagine AIXPLORER ® & A IXPLORER® Ultimate ultrasound diagnostic systems and transducers are intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging , Doppler fluid flow analysis of the human body.

The SuperSonic Imagine AIXPLORER ® & A IXPLORER® Ultimate ultrasound diagnostic systems are indicated for use in the following applications: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, Intraoperative, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac.

The system s also provide the ability to measure anatomical structures (Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Peripheral Vascular, Intraoperative, GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal, Neonatal/Adult Cephalic, Fetal/Obstetrics, Cardiac).

6) Summary of Technological Characterisitics – New Device compared to Predicates

Toshiba

SuperSonic Imagine

New devices

Toshiba Device

(predicate) SuperSonic Imagine AIXPLORER®

(predicate) SuperSonic Imagine AIXPLORER® & AIXPLORER® Ultimate

(New devices)

510(k) Number K141459 K161999 Unassigned Classification Name Ultrasonic Pulsed Doppler Imaging System (892.1550) Identical Identical Ultrasonic Pulsed Echo Imaging System (892.1560) Diagnostic Ultrasound Transducer (892.1570) Class Class II Identical Identical SUPERSONIC IMAGINE page 3 of 7

Toshiba

SuperSonic Imagine

New devices Intended Use Identical Identical Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body

General Description General purpose, mobile, software controlled diagnostic ultrasound system. To acquire ultrasound data and to display the data in various modes of operation. Identical Identical Consists of two parts: the system console and the transducer. The system console contains the user interface, a display, system electronics and optional peripherals (printers, etc…). Identical Identical Includes physical knobs and buttons of the main control panel and the user interface which consists of a Touch Panel, to access additional less-frequently-used controls, and the Alphanumeric Keyboard to enter patient data and other text. Identical Identical SWE Dynamic Range --- Adjustable range capability and numeric display with scale Identical to K161999 Clinical Applications

Abdominal, Identical Identical Small Organs* Identical Identical Musculoskeletal Identical Identical Superficial Musculoskeletal Identical Identical Fetal Identical Identical GYN Identical Identical SUPERSONIC IMAGINE page 4 of 7

Toshiba

SuperSonic Imagine

New devices Cardiac Identical Identical Adult and neonatal cephalic Identical Identical Pediatric Identical Identical Urology Identical Identical Vascular Identical Identical Peripheral Vascular Identical Identical Intra-operative Identical Identical Trans-rectal Identical Identical Trans-vaginal Identical Identical

Imaging Modes

Conventional B-mode (Harmonic/Fundamental) Identical Identical M-mode, Identical Identical PW, Identical Identical CW (continuous Wave), --- ---

Color Doppler,

Identical Identical Amplitude Doppler (CPI),

Microvascular SMI

Identical

Identical (called Angio PL.U.S) Identical

Identical (called Angio PL.U.S) Other

Spatial Compounding, Panoramic, Identical Identical Contrast Identical Identical Combination of modes Identical Identical

Design

Cart Mobile cart-based product with control panel and monitor Identical Identical Controls Typical ultrasound imaging controls including power output, gain, depth, focus, freeze, PRF, mode select Identical Identical SUPERSONIC IMAGINE page 5 of 7

Toshiba

SuperSonic Imagine

New devices Transducers

Transducer types Linear Array Identical Identical Curved Array Identical Identical Phased Array Identical Identical Motorized Linear Probe Identical Identical Microconvex probe Identical Identical Biopsy guide Yes BEAM Yes --- Yes Identical (called Needle PL.U.S.) Track Track 3 (Acoustic Output Display) Identical Identical Patient Contact Materials Yes, per ISO-10993-1 Identical Identical Acoustic Output within FDA guidelines Yes, as per NEMA UD-3 Identical Identical

Image Review Yes Identical Identical Measurement Package Yes Identical Identical Calculation Package Yes Identical Identical Report Yes Identical Identical General Safety Conforms to IEC 60601- 1, IEC 60601-1-2, IEC 60601-2-37 Identical Identical

Labeling Conforms to 21 CFR Part 801 Identical Identical General Description Consists of two parts: the system console and the transducer. The system console contains the user interface, a display, system electronics and optional peripherals (printers, etc…). Identical Identical General Description SWE Dynamic Range Includes physical knobs and buttons of the main control panel and the user interface which consists of a Touch Identical Identical SUPERSONIC IMAGINE page 6 of 7

Toshiba

SuperSonic Imagine

New devices Panel, to access additional less-frequently-used controls, and the Alphanumeric Keyboard to enter patient data and other text. --- Adjustable range capability and numeric display with scale Identical to K161999 Note: *: Breast, Thyroid, Testicle, etc

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 29, 2026