510(k)K171695

FDA DEVICE FACTS

First Sign Multi-Drug Dip Card Test, First Sign Multi-Drug Cup Test, First Sign Drug of Abuse Dip Card Test Marijuana, First Sign Drug of Abuse Cup Test Marijuana

W.H.P.M., Inc.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEFebruary 5, 2018PRODUCT CODENFT

Indications for use

Complete original indications for use part 1

First Sign® Drug of Abuse Cup Test Marijuana is a qualitative lateral flow immunoassay intended for the detection of Marijuana in human urine at cut-off concentration of 20 ng/mL. The tests provide only preliminary test results. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be used when you get any drug of abuse test result. It should be used particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for prescription use. First Sign® Drug of Abuse Dip Card Test Marijuana is a qualitative lateral flow immunoassay intended for the detection of Marijuana in human urine at cut-off concentration of 20 ng/mL. The tests provide only preliminary test results. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be used when you get any drug of abuse test result. It should be used particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for prescription use.

Complete original indications for use part 2

First Sign Multi-Drug Cup Test is a qualitative lateral flow immunoassay intended for the detection of Amphetamine, Cocaine, and Methamphetamine in human urine at cut-off concentrations of 500 ng/mL, 150 ng/mL, and 500 ng/mL, respectively. The tests provide only preliminary test results. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be used when you get any drug of abuse test result. It should be used particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter use. First Sign® Multi-Drug Dip Card Test is a qualitative lateral flow immunoassay intended for the detection of Amphetamine, Cocaine, and Methamphetamine in human urine at cut-off concentrations of 500 ng/mL, 150 ng/mL, and 500 ng/mL, respectively. The tests provide only preliminary test results. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be used when you get any drug of abuse test result. It should be used particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter use.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 28, 2026