FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEDecember 8, 2017PRODUCT CODELHQ

Indications for use

The Med-Hot Thermal Imaging Systems are intended to review, measure and record skin temperature patterns and variations emitted from the human body. They are intended for use as adjunctive diagnostic imaging for thermally significant indications in the regions of the head and neck, breast, chest, abdomen, back and extremities. The significance of the value of these thermal patterns is determined by professional investigation. This device is intended for use by qualified technical personnel trained in its use. Predicate Device Comparison: Synopsis of the comparison analysis:

• Med-Hot Thermal Imaging Systems and the predicate device (MTI 2000) are intended for use as an adjunct to other clinical diagnostic procedures in the diagnosis, quantifying, screening of differences in skin surface temperature changes.

• Both view live imagery from the camera, record snap shots to disk, include post-processing of fast thermal events, provide analysis by region of interest (spot, line, square) and save the information in a file organizer.

• The Med-Hot TotalVision Capture software is an evolution of the original Med-Hot software program included in K063047. Improvements and new features added were based on user-driven needs and requests over the years. Comparison of Intended Use and Principle of Operation: As thermal sensors, the principle operation of all thermal imaging or infrared cameras is inherently the same, regardless of technical specifications or applications. They capture and record objects within the camera view for evaluation of thermal variations. For clinical or medical application, the information from and infrared device is strictly limited to thermal findings. The location of these findings on the body, combined with intensity and distribution of thermal patterns may lead the reader/observer to intentionally or incidentally make a thermal correlation with the expected or abnormal patterns to the anatomy in that region of the body. The Med-Hot Thermal Imaging Systems is substantially equivalent in intended use, principle of operation and similar software features to the FDA cleared Med-Hot Thermal Imaging, Inc. MTI2000. Non-Clinical Performance Testing:

• ISO 14971, Medical devices - application of risk management to medical Devices

• IEC 60068-2-27 Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock

• IEC 60068-2-6 Environmental testing - Part 2-6: Tests - Test Fc: Vibration

• IEC 60529 Degrees of protection provided by enclosures (IP Code)

• EN 61000-6-3 Electromagnetic compatibility (EMC). Generic standards. Emission standard for residential, commercial and light-industrial environments

• EN 61000-6-2 Electromagnetic compatibility (EMC). Generic standards. Immunity for industrial environments

• EN 60950 Information technology equipment - Safety - Part 1: General requirements

• 47 CFR Part 15 - RADIO FREQUENCY DEVICES

• Software Verification and validation according to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Clinical Studies: No clinical studies were conducted.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 28, 2026