RARA CRAFT DATABASE for FDA Medical Devices
510(k)K172444

FDA DEVICE FACTS

ROSA BRAIN (v3.0.0.5)

Medtech S.A

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEMarch 28, 2018PRODUCT CODEHAW

Indications for use

Intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

Identical Where used Neurosurgical operating room Identical User Neurosurgeon Identical General device description Computer controlled electromechanical 6-axis multi-jointed arm

Identical Principle of operation  Preoperative images  Surgical planning  Patient registration  Guidance of instruments

Identical Preoperative images & surgical planning Preoperative images 3D MRI / CT Identical DICOM compliance DICOM 3.0 Identical Multimodality fusion Yes (MRI/CT) Identical Planning software ROSANNA BRAIN (v3.0.0.0) (product version)(Medtech) ROSANNA BRAIN (v3.0.0.5) (product version) (Medtech)

Justification: Upgraded version. Integrated planning software Yes Identical Define regions of interest (ROI) Yes Identical Trajectory definition (stereotactic module) Parameters for planning trajectories: entry point, target point, instrument length, diameter, name, color Identical Trajectory definition (endoscopy module) Parameters for planning trajectories: entry point, target point, instrument length, diameter, name, security radius (10mm by default), security aperture (10° by default) Identical Save/load planning Yes Identical Patient registration Localization means Robot arm absolute encoders Identical Device ROSA BRAIN 3.0.0.0 (K151359) ROSA BRAIN 3.0.0.5 (submission subject) Controller Axis controller for each joint Kinematic transformation between the Cartesian space and joint space Supervisor module Identical Registration methods  Fiducial markers  Optical registration device  Stereotactic frame (fiducials mounted on the frame) Identical Fiducial markers registration with pointer probe Yes Identical Optical registration with laser telemeter Yes Identical Laser class for optical registration 2 (complies with 21 CFR 1040.10) Identical Cooperative movement Yes Identical Accuracy verification on anatomical landmarks Yes (navigation probe) Identical Instruments guidance Image-guided Yes Identical Display real-time instrument position on preoperative images Yes Identical Mechanical guidance for surgical instruments Yes Identical Instrument guide position adjustment Automatic (robotized) Identical Surgeon carries out final gesture through the instrument guide with traditional surgical instrument Yes Identical Instrument fixation Instruments are mounted onto robot arm’s flange Identical Instrument calibration method Factory calibration Identical Components  Navigation probe  Standard tool holder  Endoscope holder  Microdrive holder  Optical sensor  Fiducial markers  Head holder adaptor  Leksell frame registration plates

 Navigation probe  Standard tool holder  Endoscope holder  Microdrive holder  Optical sensor  Fiducial markers  Head holder adaptor  Leksell frame registration plates  CRW Frame

Justification: Addition of a compatibility to a new head holder ("CRW Frame") by adding a mechanical component the "CRW Frame Adaptor" to immobilize the patient's head specifically during DBS surgery procedure Patient immobilization Yes - The device is attached to the head holder or the frame via an Yes – The device is attached to a head holder or a frame via an adaptor Device ROSA BRAIN 3.0.0.0 (K151359) ROSA BRAIN 3.0.0.5 (submission subject) adaptor Different - Same principle with addition of a new head holder CRW Frame Device mobility Yes - Mobile stand with wheels, immobilized with 4 stabilization feet Identical Vigilance system Yes - pedal Identical

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 28, 2026