GORE Molding and Occlusion Balloon Catheter
W. L. Gore and Associates, Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The GORE® Molding and Occlusion Balloon Catheter is intended for temporary occlusion of large vessels or to assist in the expansion of self-expanding endovascular prostheses (stent grafts).
Summary of Similarities and Differences in Technological Characteristics
Technical Equivalence is established through similar design of single lobed balloon catheter and profile, principal of operation by hydraulic, manual syringe inflation, specification of operating/treatment diameter ranges, fundamental technology and material properties, and user interface.
A comparison of the technical characteristics of the subject and predicate devices is provided in the following table. GORE® Molding and Occlusion Balloon Catheter Section 5: 510k Summary
Comparison of the GORE® Molding and Occlusion Balloon Catheter to Predicate Device (NOTE: Shaded table cells indicate characteristics that are identical)
GORE® Molding and Occlusion Balloon Catheter Q50® PLUS Stent Graft Balloon Catheter Difference Comparison Operating Principal Device is inserted and withdrawn over a guidewire and through an introducer sheath to the treatment site, via fluoroscopic guidance with the aid of balloon radiopaque marker bands. Device hub allows inflation/deflation of the balloon via a manual syringe. Device is inserted and withdrawn over a guidewire and through an introducer sheath to the treatment site, via fluoroscopic guidance with the aid of balloon radiopaque marker bands. Device hub allows inflation/deflation of the balloon via a manual syringe.
Max Inflation Volume 48ml 60ml Subject device does not exceed maximum inflation volume of predicate. Treatment Diameter 10mm – 37mm 10mm – 50mm Subject device is within the range of treatment diameter of the predicate device. Design of Construction Single-lobed balloon Single-lobed balloon Catheter shaft 3 lumen 3 lumen Working Length 90cm 65cm, 100cm Subject device is within the range of lengths offered for the predicate device. Profile 10 Fr 12 Fr Subject device does not exceed the profile of the predicate device. Guidewire Compatibility 0.035” 0.035” Radiopacity Dual balloon marker bands on catheter Dual balloon marker bands on catheter
Balloon diameter versus volume 10mm – 2ml 20mm – 7ml 30mm – 16ml 37mm – 35ml 10mm – 3ml 20mm – 6ml 30mm – 16ml 40mm – 32ml 50mm – 60ml Balloon Diameter vs Volume of the devices are similar, but not identical. Predicate device maximum balloon diameter vs volume does not exceed that of the predicate device. User Interface 2 port y-arm 2 port y-arm Balloon Materials of Construction Compliant Polyurethane Compliant Polyurethane
Catheter Materials of Construction Pellethane Pebax Catheter materials of construction are different. Sterilization Method Ethylene Oxide Ethylene Oxide GORE® Molding and Occlusion Balloon Catheter Section 5: 510k Summary