FDA DEVICE FACTS
Portable X-ray System
Rolence Enterprise, Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The XR-01 Portable X-Ray System is intended to be used by trained dentists and dental technicians as an extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and child subjects.
Performance Testing Subject device XR-01 complies with safety and performance standards which are in accordance with 21 CFR 1020.30 and 31. Performance test validate that accuracy, reproducibility, deviation and linearity of technical factors such as tube voltage, tube current and exposure time is meet the criteria. Safety tests in terms of HVL, leakage, x-ray field size meet Federal standard requirements.
Standards and Guidance Documents
Electrical Safety and EMC Standards IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 EN 60601-1-2:2015 IEC 60601-1-3:2008 (Second Edition) + A1:2013 for use in conjunction with IEC 60601-1:2005 (Third Edition) + A1:2012 IEC 60601-1-6:2010 (Third Edition) + A1:2013 IEC 60601-2-65:2012 (First Edition) for use in conjunction with IEC 60601-1:2005 (Third edition)
Battery Safety Standards IEC 62133:2012 (Second Edition)
Usability Standards IEC 62366: 2007 (First Edition) + A1: 2014
Radiation Control 21 CFR 1020.30 21 CFR 1020.31 Rolence Enterprise Inc. No. 18-3, Lane 231, Pu chung Rd., Chungli, Taoyuan, 32083 Taiwan
Guidance Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use How to Prepare a Special 510(k) Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Pediatric Information for X-ray Imaging Device Premarket Notifications Applying Human Factors and Usability Engineering to Medical Devices Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Refuse to Accept Policy for 510(k)s eCopy Program for Medical Device Submissions