Indications for use
The Amsco Evolution Series Steam Sterilizer Models HC-2000 and HC-3000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:
• Prevacuum – is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
• Steam Flush Pressure-Pulse (SFPP) – is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The Amsco Evolution Series Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 6-1): Table 6-1. Amsco Evolution Series Prevacuum Steam Sterilizer factory-programmed sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load Prevac 270°F (132°C) 4 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 6-3 for recommended quantities. Prevac 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack Gravity 250°F (121°C) 30 minutes 15 minutes Fabric Packs. Refer to Table 6-3 for recommended quantities. Liquid* 250°F (121°C) 45 minutes N/A Refer to Table 4-4 for recommended quantities. Prevac 275°F (135°C) 3 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. Gravity 270°F (132°C) 15 minutes 45 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. Gravity 250°F (121°C) 30 minutes 45 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. Gravity 270°F (132°C) 25 minutes 15 minutes Fabric Packs. Refer to Table 6-3 for recommended quantities. DART Warm-Up 270°F (132°C) 3 minutes 1 minute N/A DART 270°F (132°C) 3 ½ minutes 1 minute Bowie-Dick Test Pack, DART Test Pack Leak Test N/A N/A N/A N/A *The liquid cycle is for non-patient contact use only. The Amsco Evolution Series Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 6-2): Table 6-2. Amsco Evolution Series Steam Flush Pressure-Pulse (SFPP) Sterilizer factory-programmed sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load SFPP 270°F (132°C) 4 minutes 45 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 6-3 for recommended quantities. SFPP 275°F (135°C) 3 minutes 45 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. Prevac 270°F (132°C) 4 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 6-3 for recommended quantities. Gravity 250°F (121°C) 30 minutes 15 minutes Fabric Packs. Refer to Table 6-3 for recommended quantities. SFPP 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack Prevac 275°F (135°C) 3 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. Gravity 270°F (132°C) 15 minutes 45 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. Liquid* 250°F (121°C) 45 minutes N/A Refer to Table 6-4 for recommended quantities. DART Warm-Up 270°F (132°C) 3 minutes 1 minute N/A DART 270°F (132°C) 3 ½ minutes 1 minute Bowie-Dick Test Pack, DART Test Pack Leak Test N/A N/A N/A N/A *The liquid cycle is for non-patient contact use only. The following table lists STERIS’s recommended loads by sterilizer size: Table 6-3. Amsco Evolution Series Steam Sterilizer recommended loads per sterilizer size Sterilizer Size Wrapped Instrument Trays Fabric Packs 26” x 61” x 49” HC-2000 15 40 26” x 61” x 72” HC-3000 25 60 The following table is a guideline for liquid cycle processing: Table 6-4. Amsco Evolution Series Steam Sterilizer Liquid Cycle Guideline Number of Containers Volume of Liquid In One Container Minimum Recommended Sterilize Time at 250ºF (121ºC) 3 1000 ml 45 minutes The Amsco Evolution Series Steam Sterilizer is offered in the following chamber sizes:
• Model HC – 2000 26” x 61” x 49”
• Model HC – 3000 26” x 61” x 72”