RARA CRAFT DATABASE for FDA Medical Devices
510(k)K180754

FDA DEVICE FACTS

C-Zero Pedicle Screw System, Freedom Pedicle Screw System

Signature Orthopaedics Pty, Ltd.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEDecember 7, 2018PRODUCT CODENKB

Indications for use

The Signature Orthopaedic C-Zero Pedicle Screw and Freedom Pedicle Screw Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities and deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; and failed previous fusion (pseudarthrosis). In addition, the C-Zero Pedicle Screw and Freedom Pedicle Systems are intended for skeletally mature patients in the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra and degenerative spondylolisthesis with objective evidence of neurologic impairment. Performance Testing: Non-clinical testing and engineering evaluations were conducted to verify that the performance of the Signature Orthopaedic C-Zero Pedicle Screw System and Freedom Pedicle Screw System are adequate for anticipated in-vivo use. The following non-clinical testings were carried out on the worst case sizes of the subject pedicle screws:

• Static and dynamic compression bending testing

• Static torsion testing

• Static Flexion-Extension Moment Test

• Axial Gripping Test

• Axial Torsional Test

• Axial Gripping Test of connector components

• Torsional Gripping Test of connector components

• Screw insertion testing

• Screw pull-out testing

• Screw torque to failure testing

• Freedom Dissociation Test at the screw-tulip interface Testing Standards:

• ASTM-F1717_Standard Test Methods for Spinal Implant Construct in a Vertebrectomy Model

• ASTM-F1798-13_Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants

• ASTM-F543_Standard Specification and Test Method for Metallic Medical Bone Screw

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 26, 2026