FDA DEVICE FACTS
Trilogy Evo
Respironics, Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The Trilogy Evo ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional, home, and non-emergency transport settings for example wheelchair or personal vehicle. It may be used for both invasive and non-invasive ventilation.
The Intended Use of the Trilogy Evo is the same as compared to the primary predicate, Trilogy Series of Ventilator with Oximetry, K111610, to provide invasive and noninvasive positive pressure ventilation. The Indications for Use for the Trilogy Evo is not identical to the predicate device, Trilogy Series of Ventilator with Oximetry, K111610; however, the differences do not alter the intended therapeutic use of the devi ce nor do they affect the safety and effectiveness of the device relative to the predicate.
Comparison of Technological Characteristics with the Predicate Device The Trilogy Evo ventilator is similar to the Primary Predicate Device, Trilogy Series of Ventilator with Oximetry, K111610. The Trilogy Evo ventilator has the same intended use and similar indications for use, operating principles, technologies and manufacturing processes as the predicate device. See table below for a comparison of the Trilogy Evo ventilator to the primary predicate device, Trilogy Series of Ventilator with Oximetry, K111610. Trilogy Evo Ventilator Traditional 510(k) Respironics Inc. Page 6-4 Comparison of Technological Characteristics with the Predicate Device
Characteristic Subject Device: Trilogy Evo Manufacturer: Respironics
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。