FDA DEVICE FACTS
Cardinal Health Sterilization Wrap
Cardinal Health200, LLC
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
Cardinal Health™ Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by using:
• Pre-vacuum steam at 270°F/132°C for 4 minutes
• Gravity steam at 250°F/121°C for 30 minutes
• 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes
• Advanced Sterilization Products (ASP) STERRAD® 100S System
• Advanced Sterilization Products (ASP) STERRAD® 200 System
• Advanced Sterilization Products (ASP) STERRAD® NX System, Standard and Advanced Cycles
• Advanced Sterilization Products (ASP) STERRAD® 100NX, Standard, Flex, Express, and DUO cycles
• Lumen, Non-Lumen, and Flexible Cycles in the STERIS V-PRO® 1, STERIS V- PRO® 1 Plus, STERIS V-PRO® maX and STERIS V-PRO® 60 Low Temperature