RARA CRAFT DATABASE for FDA Medical Devices
510(k)K181871

FDA DEVICE FACTS

EliA Celikey IgG Immunoassay; EliA GliadinDP IgA Immunoassay; EliA GliadinDP IgG Immunoassay

Phadia AB

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEMarch 1, 2019PRODUCT CODEMVM

Indications for use

EliA GliadinDP IgG EliA GliadinDP IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to gliadin in human serum and plasma (heparin, EDTA, citrate) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA GliadinDP IgG uses the EliA IgG method on the instrument Phadia 250. EliA GliadinDP IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA GliadinDP IgG uses the EliA IgG method on the instrument Phadia 2500/5000. 006_A.6 510k Decision Summary Input - V6 A.6-6 Analytical technology: Immuno-fluorescence measurement Same Same Assay process Same Same Common, dedicated Phadia reagents Same Same Introduction of new article numbers for Development Solution, Stop Solution and Washing Solution is only due to larger filling volumes which are required for the bigger instruments Phadia 2500/5000 Result calculation software; Phadia Information Data Manager (IDM) Same Same Sample volume 90 µL (20 µL of non-diluted sample) 90 µL (20 µL of non-diluted sample) Incubation temperature 37°C 37°C Conjugate volume 90 µL 90 µL Development Solution Volume 90 µL 90 µL Stop Solution Volume 200 µL 200 µL Assay set-up Random access Random access Reagent packaging size Various/Common Various/Common Introduction of new article number for EliA Sample Diluent (83-1071-01) is only due to larger filling volume. Onboard storage of reagents Yes Yes Time to 1st result ~2 h ~2 h 006_A.6 510k Decision Summary Input - V6 A.6-7 EliA Celiac Immunoassays on Phadia 250 and Phadia 2500/5000 instruments – Differences to predicate device

Feature Predicate Device Phadia 250 New Device Phadia 2500/5000 Sample matrix; Serum or plasma type as indicated in the DFU dependent on assay For EliA Celikey IgG: Serum or plasma (EDTA, citrate) For EliA GliadinDP IgA and IgG: Serum or plasma (heparin, EDTA, citrate) For EliA Celikey IgG: serum or EDTA-plasma For EliA GliadinDP IgA and IgG: Serum or plasma (Li-heparin, EDTA) Daily throughput ~250 tests ~2500/5000 tests Sample Dilution Phadia 250 uses a steel pipette to dilute the samples in Dilution Plates (Art.No. 12-3907-08) Phadia 2500/5000 uses disposable Pipette Tips in Racks (Art No. 12-3805-04) for pipetting samples in Dilution Well (Art.No. 12-4005-69) Risk for carry-over The warning “DO NOT REUSE” in the Phadia 250 DFU for EliA Conjugates is due to the fact that a low risk of conjugate contamination by carry-over from samples was identified. In order to reduce the risk, the single use statement for the conjugate was included in the Phadia 250 DFU. When running EliA tests on the Phadia 2500/5000 instruments, there is no need for this warning statement because these instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, and carry-over from samples to conjugate is impossible. Loading of EliA Carriers EliA carriers are loaded manually on the Loading Tray from where they can be processed directly or transferred to the cooled storage compartment. The Phadia 2500/5000 instruments do not have such a Loading Tray. The EliA carriers are loaded into racks which are directly transferred to the cooled storage compartment Barcode reader The Phadia 250 instrument has a built-in barcode reader at the front of the instrument, but the operator needs to scan the barcodes manually by showing the reagents to the barcode reader. Alternatively, the operator can also enter the characters below the barcode manually. The Phadia 2500/5000 instruments dispose of a built-in barcode reader, and the reagents are on a moving belt which conveys them past the barcode reader. The lot-specific information will be read automatically by the instrument during loading. Process time / Time to patient result Phadia 250 needs 1 minute to process one Well. Phadia 250 provides the results at a one minute interval. Phadia 2500/5000 instruments process two Wells in parallel in 48 seconds. Phadia 2500/5000 provides the results at a 24 seconds interval. 006_A.6 510k Decision Summary Input - V6 A.6-8

K. Standard/Guidance Document Referenced (if applicable): CLSI EP05-A3; Evaluation of Precision Performance of Quantitative Measurement Methods; September 201 4 CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; April 2003 CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantification; October 2004. CLSI EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 25, 2026