Cook Unimpregnated Central Venous Catheter
Cook Incorporated
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
- Function of catheter: Both the predicate and subject devices are indicated for continuous or intermittent drug infusions, central venous blood pressure monitoring (CVP), acute hyperalimentation, blood sampling, delivery of whole blood or blood products, and power injection of contrast media. The subject device is also indicated for simultaneous, separate infusion of drugs (for multi-lumen catheters only). Since a multi-lumen CVC allows simultaneous and separate infusion of drugs by utilizing individual lumens, this additional language in the indications for use statement does not raise different questions of safety and/or effectiveness when compared to the predicate device.
- Power injection of contrast media: All French sizes of the predicate device catheter are indicated for power injection of contrast media, and the flow rate may not exceed 10 mL/sec. The subject device includes catheters that are indicated for power injection of contrast media and catheters that are not indicated for power injection of contrast media. For the power injectable catheters, the flow rate may not exceed 3.0 mL/sec for the 4.0 and 5.0 French catheters and 10 mL/sec for the 7.0, 8.0, 9.0, and 10.0 French catheters. These flow rate limits were established from performance testing using subject device catheters with power injector pressures set at the maximum safety cut-off pressure to ensure safe and effective use of the catheters. For the 4.0 and 5.0 French subject device catheters, the Cook Incorporated – Traditional 510(k) 12 Cook Unimpregnated Central Venous Catheter May 16, 2019
maximum flow rate (3.0 mL/sec) is within the range of a reference device, the Turbo-Ject Peripherally Inserted Central Venous Catheter (PICC) (K072625), which is available in 4.0 and 5.0 French with maximum flow rates ranging from 3 mL/sec to 7.0 mL/sec. For the 7.0, 8.0, 9.0, and 10.0 French subject device catheters, the maximum flow rate (10 mL/sec) is identical to the predicate. The 2.5, 3.0, 6.3, and 7.5 French subject device catheters are not indicated for power injection of contrast media. The non-power injectable subject device catheters, not having an indication for power injection of contrast media does not raise different questions of safety and/or effectiveness when comparing to the predicate. In addition, the subject device will include two Instructions for Use (IFU) booklets; one for the power injectable catheters and one for the non-power injectable catheters. For the non-power injectable catheters, the IFU states the following in the Warning Section: “Do not power inject contrast medium through catheter. Catheter rupture may result.”
- Patient Population: The predicate device does not specify age / weight recommendation for patient use. The subject device is indicated for adult and pediatric patients. Central venous catheters are critical in establishing a vascular access to sample blood, monitor blood pressure, and administer fluids intravenously, and these functions apply to both adult and pediatric patients. It is common in clinical practice to use relatively small CVCs in pediatric patients because their blood vessels are generally smaller comparing to adult patients. When establishing the minimum age and weight of patients for each catheter French size, the approximate diameters of femoral veins from a wide range of pediatric patients (from preterm neonates to 19 years old) have been carefully considered to ensure each French size is appropriate for use in patients with the associated minimum age/weight. Overall, the subject and predicate devices have the same intended use and similar indications for use.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。