DECISION DATEFebruary 26, 2019PRODUCT CODENGX

Indications for use

The Body System is intended for muscle conditioning to stimulate healthy muscles. The Body System is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. The Body System is intended to be operated by a trained professional who is present to monitor treatment. Predicate Device: The Body System is substantially equivalent to the predicate devices listed below:

• K123158, Ion Magnum Genius, under 21 CFR 890.5850 and Product Code NGX

Predicate Trade Name: Ion Magnum Genius Body System Traditional 510(k) Revision 4.0 Premarket Notification Page 9-3 Classification Name: Stimulator, Muscle, Powered for Muscle Conditioning Premarket Notification: K123158 Manufacturer: Ion Genius, Inc.

• K143551, Compex Wireless USA, under 21 CFR 890.5850 and Product Code NGX

Predicate Trade Name: Compex Wireless USA Classification Name: Stimulator, Muscle, Powered for Muscle Conditioning Premarket Notification: K143551 Manufacturer: DJO, LLC

Based on the intended use/Indications for use, design, materials, principle of operation, energy used and materials, the Body System is substantially equivalent to the legally marketed predicate devices. The non-clinical performance data demonstrates substantial equivalence and includes an analysis of output waveforms. Non-Clinical Tests Performed on the Device: The Body System has been tested to establish is safety and performance. The following tests have been performed on the patient contacting parts:

• Biocompatibility – Cytotoxicity (ISO-10993-5) – result: Pass

• Biocompatibility – irritation (ISO-10993-10) – result: Pass

• Biocompatibility – Sensitization (ISO-10993-10) – result: Pass

• Performance – including Output Waveforms of the Body System as compared to a predicate

• Electromagnetic Compatibility – according to EN 60601-1-2:2015, IEC 60601-1-2:2014, IEC 60601-2-10 2012 + A1 2016

• Electrical Safety – according to IEC 60601-1-6:2010, AMD1:2013, IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) Summary: There are no differences in the technological characteristics between the subject device and the predicate devices, which raise issues concerning safety, performance or efficacy. The Body System is substantially equivalent to the predicate device(s) in terms of safety and effectiveness. Body System Traditional 510(k) Revision 4.0 Premarket Notification Page 9-4 Comparison of the Subject and Predicate Devices Elements of Comparison Subject device Primary

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 25, 2026