510(k)K182865

FDA DEVICE FACTS

Neodent Instruments Kits

Jjgc Industria E Comercio DE Materiais Dentarios S.A.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEMay 7, 2019PRODUCT CODEKCT

Indications for use

Complete original indications for use

Indications for Use for GM/WS Surgical Kit Case: Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kits require the use of FDA cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 20 minute dry time Gravity displacement – Exposure at 132 °C for 15 minutes, 20 minute dry time Neodent Instrument Kits are intended for sterilization of non-porous loads. The combined weight of the GM/WS Surgical Kit Case and the associated instruments is 674.5 g. The weight of the empty Kit Case is 507 grams. Neodent Instrument Kits are recommended not to be stacked during sterilization.

Indications for Use for GM Prosthetic Kit Case: Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kits require the use of FDA‐cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA‐cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum (pre‐vacuum) – Exposure at 132 °C for 4 minutes, 20‐minute dry time Gravity displacement – Exposure at 132 °C for 15 minutes, 20‐minute dry time. Neodent Instrument Kits are intended for sterilization of non‐porous loads. The The combined weight of the GM Prosthetic Kit Case and the associated instruments is 250.5 g. The weight of the empty Kit Case is 210 grams. Neodent Instrument Kits are recommended not to be stacked during sterilization.

Indications for Use for GM Try‐In Kit Case: Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care pro vider Neodent Instrument Kits require the use of FDA‐cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilizable wrap that is FDA‐cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum (pre‐vacuum) – Exposure at 132 °C for 4 minutes, 20‐minute dry time. Gravity displacement – Exposure at 132 ºC for 15 minutes, 20‐minute dry time. Neodent Instrument Kits are intended for sterilization of non‐porous loads. The combined weight of the GM Try -In Kit Case and the associated instruments is 212.6 g. The weight of the empty Kit Case is 195 grams. Neodent Instrument Kits are recommended not to be stacked during sterilization.

Indications for Use for GM Guided Surgery Kit Case: Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kits require the use of FDA‐cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA‐cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum (pre‐vacuum) – Exposure at 132 °C for 4 minutes, 20‐minute dry time Gravity

displacement – Exposure at 132 °C for 15 minutes, 40‐minute dry time. Neodent Instrument Kits are intended for sterilization of non‐porous loads. The combined weight of the GM Guided Surgery Surgical Kit Case and the associated instruments is 728.4 g. The weight of the empty Kit Case is 567 grams. Neodent Instrument Kits are recommended not to be stacked during sterilization.

See the full product information製品情報をすべて見るには