FDA DEVICE FACTS
BONEBRIDGE System, Bone Conduction Implant Kit (BCI 602 Implant Kit), BCI 602 Sizer Kit, BCI 602 Lifts (1mm)
Med-El Elektromedizinische Geraete GmbH
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
INTENDED USE:
The Bonebridge is intended to improve hearing for paticnts with conductive or mixcd hearing losses, bilatcral fitting, and single-sided deafness.
INDICATIONS:
The Bonebridge bone conduction hearing implant system is intended for the following patients and indications:
- Patients 12 years of age or older.
- Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) should be better than or equal to 45 dB HL.
- Bilateral fitting of the Bonebridge is intended for patients having a symmetrically conductive or mixed hearing loss. ‘The difference between the left and right sides’ BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 3 kHz, or less than 15 dB at individual frequencies.
- Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite car
(i.e., single-sided deafness or ““SSD”). The pure tone average air conduction hearing thresholds of the
hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
-The Bonebridge for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
- Prior to receiving the device, it is recommended that an individual has experience with appropriately fit air conduction or bone conduction hearing aids.
CONTRAINDICATIONS:
- Chronic or non-revisable vestibular or balance disorders.
- Abnormally progressive hearing loss.
- Evidence of conditions that would prevent good speech recognition potential as determined by good clinical judgment.
- Skin or scalp conditions that may preclude attachment of the audio processor or that may interfere with the use of the audio processor.
-Skull size or abnormality that would preclude appropriate placement of the Bonebridge implant as
determined by CT scan.
Documented in IU18-019 (see attached).
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。