FDA DEVICE FACTS
EZ-TRAX Containment Device
K1 Medical, LLC
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The EZ-TRAX™ Containment Device is intended for use as an accessory in healthcare facilities to organize, enclose, reprocess, transport, and store medical devices between surgical and other medical uses. The EZ- TRAX™ Containment Device is not intended on its own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterile barrier system.
Validated Cycle Times for Dynamic Air Removal Steam Sterilization Cycles Cycle Temperature Exposure Time Drying Time Dynamic Air Removal 270° F/ 132° C 4 minutes 10 minutes
Sterilization validations included a worst case EZ-TRAX™ Containment Device and a medical device challenge set comprising of:
- Lumen dimensions (3) 1mm x 500mm
- Conjoined/mated surfaces: forceps, clamps, bending pliers, ratchet handles, retractors
- Cannulated: drill bits, taps, guides, screwdrivers, cannulated screws
- A total weight of 42 lbs comprising of (EZ-TRAXTM Containment Device + sterile barrier system + medical device load) as a worst case challenge to the system. Healthcare facilities should not exceed 25 pounds (EZ-TRAXTM Containment Device + sterile barrier system + medical device load).
The EZ-TRAXTM is offered in the following sizes:
Brand Name Model Dimensions EZ-TRAXTM BASE.ASSY.AL.24.12.2 (L) 22.97” (W) 11.18” (H) 2.04” EZ-TRAXTM BASE.ASSY.AL.24.12.2.4 (L) 22.97” (W) 11.18” (H) 2.44”
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。