510(k)K192657

FDA DEVICE FACTS

Sterile High-pressure Angiographic Syringes for Single-use

Shenzhen Boon Medical Supply Co., Ltd.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEAugust 7, 2020PRODUCT CODEDXT

Indications for use

The proposed device, sterile High-pressure Angiographic Syringes for Single -use are intended for the injection of Contrast media or saline; they shall be used with an US legally marketed angiographic injectors. Sterile High -pressure Angiographic Syringes for Single-use are intended for the injection of c ontrast media or saline; they shall be used with an US legally marketed angiographic injectors. Same Prescription only (Rx) Prescription only (Rx) Prescription only (Rx) Same Mode of operation Power-driven operation, single use Power-driven operation, si ngle use Same Configuration Angiographic Syringe Angiographic Syringe Same Connecting tube Connecting tube Same J shape tube/Spike J shape tube/Spike Same Sterility EO Sterilized EO Sterilized Same Single Use Yes Yes Same Maximum withstanding pressure Syringe 300psi, 400psi, 1200psi 300psi, 1200psi Difference #1 Connection tube 300psi, 400psi, 1200psi 300psi, 1200psi Difference #1 J shape tube / / Same Spike 400 psi / Difference #1 Specification Syringe (Volume, ml) 200, 150, 125, 130, 100, 200/200, 60/100, 125/125, 100/100, 65/65, 65/115, 60/60, 50/50 200, 150, 130, 200/200, 60/100, 65/65, 65/115, 60/60, 50/50, Difference #2 Connection tube (overall Length, mm) 200~2500, 1500~2500, 1500, 1800, 2000, 2500, 500, 750, 1000, 1200 1500, 1800, 500, 75 0, 1000, 1200 Difference #2 J shape tube (overall Length, mm) 240 240 Same Spike (overall Length, 58.8, 47.3, 1000, 2800, 1200, 2900, 180, 260, 340, 420, 500, 58.85, 47.35 Difference #2 mm) 450, 550, 600 Performance Syringe ISO 7886 ISO 7886 Same Luer connector ISO 594-1; ISO 594-2 ISO 594-1; ISO 594-2 Same Compatibility Pass Pass Same Patient-Contact Material Syringe Barrel PP ( polypropylene) or PET (Polyethylene terephthalate) PP ( polypropylene) or PET (Polyethylene terephthalate) Same Piston Polyisoprene rubber Polyisoprene rubber Same Lubricant Polydimethylsiloxane Polydimethylsiloxane Same Connection tube Tubing PVC (Polyvinylchloride with DEHP) or PVC (Polyvinylchloride without DEHP) or PU (Polyurethane) PVC ( Polyvinylchloride) or PU (Polyurethane) Difference #3 Luer connectors PC (Polycarbonate) PC (Polycarbonate) Same UV adhesive Ultraviolet adhesive Ultraviolet adhesive Same Spike Closure-piercing device ABS (acrylonitrile-butadiene-styrene) ABS (acrylonitrile-butadiene-styrene) Same Filter membrane PP (polypropylene) PP (polypropylene) Same Tube Polyvinyl chloride (PVC) / Difference #3 Luer Connector Polycarbonate (PC) Protective cap Acrylonitrile-butadiene-styrene (ABS) J shape tube tube PE (Polyethylene) PE (Polyethylene) Same Biocompatibility

No Cytotoxicity No Cytotoxicity Same No Irritation No Irritation Same No Sensitization No Sensitization Same No Pyrogen No Pyrogen Same No Acute Toxicity No Acute Toxicity Same No Hemolysis No Hemolysis Same Endotoxin Limit 20 EU per device 20 EU per device Same Shelf Life 5 year 3 year Difference #4

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。