RARA CRAFT DATABASE for FDA Medical Devices
510(k)K200253

FDA DEVICE FACTS

Megadyne Telescoping Smoke Evacuation Pencil - 10ft tubing, Megadyne Telescoping Smoke Evacuation Pencil - 15ft tubing, Megadyne Telescoping Smoke Evacuation Soft Tissue Dissector, Megadyne Telescoping Soft Tissue Dissector

Ethicon Endo-Surgery

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEMay 27, 2020PRODUCT CODEGEI

Indications for use

MEGADYNE™ Telescoping Soft Tissue Dissectors are monopolar devices designed for general electrosurgical applications including cutting and coagulation (coag). The device s conduct an electrosurgical current from an electrosurgical unit (ESU) and deliver it to the target tissue to achieve the desired surgi cal effect. The electrosurgical active electrodes are intended to conduct radio frequency (RF) current for cutting and coagulation from the ESU to target soft tissue in a broad range of surgical procedures requiring the use of electrosurgery for cutting and cauterization. The electrosurgical active electrodes have a specific geometry that minimizes blanching and thermal damage in skin incisions when used in conjunction with the MEGADYNE™ ESU’s Advanced Cutting Effect (ACE) mode or the MEGADYNE™ ESU’s Geomet ric Electron Modulation (GEM) mode.

MEGADYNE™ Telescoping Smoke Evacuation Soft Tissue Dissector s and MEGADYNE Telescoping Smoke Evacuation Pencil s are monopolar devices designed for general electrosurgical applications including cutting and coagulation (coag) and for removing smoke generated by electrosurgery when used in conjunction with a smoke evacuation system. The devices conduct an electrosurgical current from an electrosurgical unit (ESU) and deliver it to the target tissue to achieve the desired surgical effect. The Megadyne PTFE coated electrosurgical active electrodes are intended to conduct radio frequency (RF) current for cutting and coagulation from the ESU to target soft tissue in a broad range of surgical procedures requiring the use of electrosurgery for cutting and cauterization. The MEGADYNE™ Telescoping Smoke Evacuation Soft Tissue Dissector electrosurgical active electrodes have a specific geometry that minimizes blanching and thermal damage in skin incisions when used in conjunction with the MEGADYNE™ ESU’s Advanced Cutting Effect (ACE) mode or the MEGADYNE™ ESU’s Geometric Electron Modulation (GEM) mode.

Summary of Similarities and Differences in Technological Characteristics The Megadyne Telescoping Smoke Evacuation Pencil s, Megadyne Telescoping Smoke Evacuation Soft Tissue Dissectors and Megadyne Telescoping Soft Tissue Dissectors are the same as the Predicate Device in the technological principle because they are all monopolar electrosurgical pencils used in general electrosurgical applications including cutting and coagulation of tissue. At a high level, the Subject and Predicate Devices are based on the following same technological elements:

• Supplied power by an external generator.

• Devices have the same rated voltage 5.5 kV peak.

• Same general steps for use.

• Operation function switches, used to switch between “CUT” and “COAG”

• Polycarbonate Nozzle and design

• Optional use of smoke evacuation system, on the Megadyne Telescoping Smoke Evacuation Pencil and Megadyne Telescoping Smoke Evacuation Soft Tissue Dissector

• Provided Sterile, Sterility Assurance Level SAL 10-6 The Subject Devices have a telescoping end -effector, whereas the Predicate Devices do not have a telescoping end -effector feature. The Subject Devices have a new collet material to hold the electrode. The only difference between the Predicate and Subject conductive pathway is the collet material. Although different, the resistance difference is negligible compared to the anticipated resistances observed in electrosurgical use. The Subject Devices include either an ACE Electrode or an EZ Clean Electrode. Two of the six Subject Devices, t he Megadyne™ Telescoping Soft Tissue Dissector , does not include the smoke evacuation feature to provide a non- smoke evacuation options to the surgeon, whereas the Predicate Devices have the smoke evacuation feature.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 24, 2026