DECISION DATEApril 29, 2020PRODUCT CODELNH

Indications for use

Evry is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces axial, sagittal, coronal, and oblique cross-sectional images and displays the internal structure and/or function of the head. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis. Design Features/ Fundamental Scientific Technology: Evry is based on the fundamental principle that certain atomic nuclei present in the human body can emit a weak signal when placed in a strong magnetic field. To stimulate this signal the nuclei must be excited by a radio signal at the precession frequency of the nuclei. This resonant frequency depends on the nuclei and the strength of the applied magnetic field. Emitted signals are weak radio signals at this resonant frequency which can be sampled using one or more receiver antennas and digitized for processing. Integrated into the Magnetic Resonance Imaging system are a set of gradient coils which vary the strength of the applied magnetic field as a function of position and time, and are used to spatially encode the radio signals. Computed image reconstruction algorithms are then able to convert the encoded, digitized signals into images which can then be displayed to the user. As with the predicate device, Evry includes the following technological components:

 Magnet scanner unit, gradient system and RF transmit coil  Equipment room containing the equipment needed to support the scanner unit functionality  Patient Transporter  16-Channel Receive-Only Head Coil  Patient Communication System  Operator Console Traditional 510(k) Evry TM

Summary of Non-Clinical

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 24, 2026