pREBOA-PRO Catheter
Prytime Medical Devices, Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The pREBOA-PRO™ Catheter is intended for temporary occlusion of large vessels and blood pressure monitoring including patients requiring emergency control of hemorrhage.
Comparison of Technological Characteristics with the Predicate Device
The pREBOA-PROTM Catheter is identical to the predicate device in terms of intended use and basic technological characteristics. The pREBOA-PROTM Catheter employs the same basic technologies as the identified predicate including:
• Large Vessel Occlusion with inflatable / deflatable occlusion balloon
• Fluid column pressure monitoring
The pREBOA-PRO™ Catheter has the following differences from the predicate device:
• Guidewire compatibility, including a corresponding P-tip™ change
• Semi-compliant balloon with in-line Pressure Relief (safety) Valve
• Smaller maximum inflation diameter and larger maximum inflation volume
• Longer mounted balloon length medical devices. 3 of 4
• Improved flow characteristics
• Two-sided outer shaft markings with indication of average Aortic Zone depth
Performance bench and in vivo (animal) testing was performed to support the safety and effectiveness of these modifications. The results of these tests demonstrate that the pREBOA- PRO™ Catheter has been designed and tested to conform to its intended use and is comparable to the predicate device. The modifications do not raise different questions of safety and effectiveness, and can be evaluated with performance testing. As such, it claims substantial equivalence to the predicate device.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。