ROSA ONE Brain application
Medtech S.A
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is in dicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate. The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate. Identical with the addition of explicit information on the level of training of the surgeons using the device Where used Neurosurgical operating room Neurosurgical operating room Identical User Neurosurgeon Neurosurgeon Identical Anatomical site Head Head Identical
Principle of operation
• Preoperative images
• Surgical planning
• Patient registration
• Guidance of instruments
• Pre & intraoperative images
• Surgical planning
• Patient registration
• Guidance of instruments
Identical Preoperative images & surgical planning Images type 3D MRI / CT 3D MRI / CT Identical DICOM compliance Yes Yes Identical Merge images (multimodality image fusion capability) Yes Yes Identical K200511,
Device
ROSA ONE Brain application 3.1.0.4 (K182417) ROSA ONE 3.1.3.2 Brain application (K200511)
Comparison Analysis Integrated planning software ROSANNA BRAIN (Medtech) ROSANNA BRAIN (Medtech) Identical Define regions of interest (ROI) Yes Yes Identical Trajectory planning parameters Entry point, target point, length of the instrument, diameter, name, color Entry point, target point, length of the instrument, diameter, name, color Identical
Trajectory definition (endoscopy module) Parameters for planning trajectories: entry point, target point, length of the instrument, diameter, name, security radius (10mm by default), security aperture (10° by default) Parameters for planning trajectories: entry point, target point, length of the instrument, diameter, name, security radius (10mm by default), security aperture (10° by default)
Identical Save/load planning Yes Yes Identical Patient Registration Localization means Robot arm absolute encoders Robot arm absolute encoders Identical
Controller Axis controller for each joint Kinematic transformation between the Cartesian space and joint space Supervisor module Axis controller for each joint Kinematic transformation between the Cartesian space and joint space Supervisor module
Identical
Patient registration methods
• Fiducial markers
• Optical registration device
• Stereotactic frame (fiducials mounted on the frame)
• Fiducial markers (skin, bone)
• Optical registration device
• Stereotactic frame (fiducials mounted on the frame)
Identical Fiducial markers registration with pointer probe Yes Yes Identical Surface matching registration with optical distance sensor Yes Yes Identical Laser class for optical registration Class 2 laser Wavelength – 658 nm, Maximum output
– 1 mW (complies with 21 CFR 1040.10) Class 2 laser Wavelength – 658, Maximum output – 1 mW (complies with 21 CFR 1040.10)
Identical Cooperative movement Yes Yes Identical K200511,
Device
ROSA ONE Brain application 3.1.0.4 (K182417) ROSA ONE 3.1.3.2 Brain application (K200511)
Comparison Analysis Accuracy verification on anatomical landmarks Yes Yes Identical Instruments guidance Image-guided Yes Yes Identical Real time display of the instrument position Yes Yes Identical Provide guidance for surgical instruments Yes Yes Identical Instrument guide position adjustment Automatic (robotized) Automatic (robotized) Identical Surgeon carries out final gesture through the instrument guide with traditional surgical instrument
Yes – through the instrument guide
Yes – through the instrument guide
Identical Instrument fixation Instruments are mounted onto robot arm’s flange Instruments are mounted onto robot arm’s flange Identical Instruments Instrument holder, endoscope holder and adaptors, optical sensor Instrument holder, endoscope holder and adaptors, optical sensor Identical Instrument calibration method Factory calibration Factory calibration Identical
Associated equipment
• Navigation probe
• Standard tool holder
• Endoscope holder
• Microdrive holder
• Optical sensor
• Fiducial markers
• Head holder adaptor
• Leksell frame registration plates
• CRW Frame
• Navigation probe
• Standard tool holder
• Endoscope holder
• Microdrive holder
• Optical sensor
• Fiducial markers
• Head holder
• Leksell frame registration plates
• CRW Frame Identical Patient immobilization Yes - The device is attached to the head holder or the frame via an adaptor. Yes - The device is attached to the head holder or the frame via an adaptor Identical
Device mobility Yes - Mobile stand with wheels, immobilized with 4 stabilization feet Yes - Mobile stands with wheels; Robot stand immobilized with stabilization feet
Identical Vigilance system Yes – foot pedal Yes – foot pedal Identical
Sterility Non-sterile and sterile instruments Disposable sterile drapes for the robot arm and touch screen Non-sterile and sterile instruments Disposable sterile drapes for the robot arm and touch screen
Identical K200511,
Device
ROSA ONE Brain application 3.1.0.4 (K182417) ROSA ONE 3.1.3.2 Brain application (K200511)
Comparison Analysis
Device mobility Yes - Mobile stand with wheels, immobilized with 4 stabilization feet Yes - Mobile stands with wheels; Robot stand immobilized with stabilization feet
Identical Vigilance system Yes – foot pedal Yes – foot pedal Identical
Sterility Non-sterile and sterile instruments Disposable sterile drapes for the robot arm and touch screen Non-sterile and sterile instruments Disposable sterile drapes for the robot arm and touch screen
Identical
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。