FDA DEVICE FACTS
Powder Free White Sterilized Polyisoprene Surgical Gloves-Tested for Use with The Chemotherapy Drugs
Gx Corporation Sdn Bhd
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
A Powder Free White Sterilized Polyisoprene Surgical Gloves- Tested for Use with Chemotherapy Drugs is a disposable device intended for invasive surgery procedures that is to be worn on both the hands by operating room personnel to protect a surgical wound from contamination .
These gloves were tested for use with Chemotherapy Drug as per ASTM D 6978 - Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drug.
List of tested Chemotherapy Drugs are as follows:
Chemotherapy Drug Permeation and Concentration Average Breaktrough Detection Time (in Minutes)
1. Carmustine ( BCNU)3.3 mg/ml(3,300 ppm) 12.6 min
2. Cisplatin 1.0 mg/ml (1,000ppm) >240 min
3. Cyclophosphamide (Cytoxan) 20mg/m1(20,000ppm) >240 min
4. Dacarbazine (DTIC)10.0 mg/ml1(10,000ppm) >240 min
5. Doxorubicin Hydrochloride 2.0 mg/ml (2,000ppm) >240 min
6. Etoposide (Toposar) 20.0 mg/ml (20,000ppm) >240 min
7. Fluorouracil 50.0 mg/ml (50,000ppm) >240 min
8. Ifosfamide 50.0 mg/ml (50,000ppm) >240 min
9. Methotrexate 25 mg/ml (25,000ppm) >240 min
10. Mitomycin C 0.5 mg/ml (500ppm) >240 min
11. Mitoxantrone 2.0 mg/ml (2,000ppm) >240 min
12.Paclitaxel (Taxol) 6.0 mg/ml (6,000ppm) >240 min
13. Thio-Tepa 10.0 mg/ml (10,000ppm) 12.0 min
14. Vincristine Sulfate 1.0 mg/ml (1,000ppm) >240 min
Please Note that Carmustine and Thio-Tepa have low permeation time of less than 240 minutes.