FDA DEVICE FACTS
Aerus Medical Guardian,model F170A
Aerus Medical, LLC
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The Aerus Medical Guardian, model F170A is a device intended for medical purposes that is used for the reduction of staphylococcus epidermidis and erwinia herbicola bacteria, MS2 and Phi-X174 viruses and aspergillus niger fungal spores and bacillus globigii bacterial spores from the air in a temperature-controlled professional healthcare environment of 70~71°F, 40~45% RH. The Aerus Medical Guardian, model F170A demonstrated the reduction of staphylococcus epidermidis and erwinia herbicola bacteria, MS2 and Phi-X174 viruses and aspergillus niger fungal spores and bacillus globigii bacterial spores under the following conditions. Organism Type Organism Name Test Temp/RH Exposure Time (m) Avg Log-Reduction Bacteria Staphylococcus epidermidis 72°F / 50% 60 5.95 Bacteria Erwinia herbicola 72°F / 50% 60 5.12 Virus MS2 72°F / 50% 60 5.58 Virus Phi-X174 72°F / 50% 60 4.19 Fungal Spore Aspergillus niger 72°F / 50% 60 4.12 Bacterial Spore Bacillus globigii 72°F /50% 60 4.22
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。