Getinge GSS610N Series Steam Sterilizer

Maquet GmbH

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEMarch 1, 2021PRODUCT CODEFLE

Indications for use

The Getinge GSS610N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS610N Series Steam Sterilizer is available in 3 models differentiated by chamber length; GSS610N Model 610N10 (40.9 inch chamber), GSS610N Model 610N14 (51.1 inch chamber), and GSS610N Model 610N15 (60.6 inch chamber). Cycles and Load Chart Getinge GSS610N Series Steam Sterilizer Cycles and Load Chart Cycle Type Factory Settings Load Configuration (Note 1) Maximum Items per Chamber Length Exp. Temp. Exp. Time Drying Time 610N10 1040 mm 40.9 in 610N14 1400 mm 51.1 in 610N15 1540 mm 60.6 in P1 PREVAC 1 135.0°C (275.0°F) 3 min 16 min Full instrument packs 18 24 24 P2 PREVAC 2 135.0°C (275.0°F) 3 min 3 min Full fabric packs 40 40 60 P3 PREVAC 4 132.2°C (270.0°F) 4 min 30 min Full instrument packs 18 24 24 P4 PREVAC 5 132.2°C (270.0°F) 4 min 5 min Full fabric packs 40 40 60 P5 B & D TEST 134°C (273.0°F) 3 min, 30 sec 0 min 1 B&D Test Pack in an EMPTY chamber (other than loading accessories) 1 Test Pack 1 Test Pack 1 Test Pack P6 GRAVITY 1 121.1°C (250.0°F) 30 min 45 min Full instrument packs 18 24 24 Full fabric packs 40 40 60 P7 GRAVITY 2 135.0°C (275.0°F) 10 min 45 min Full instrument packs 18 24 24 Full fabric packs 40 40 60 P8 GRAVITY 3 132.2°C (270.0°F) 15 min 45 min Full instrument packs 18 24 24 Full fabric packs 40 40 60 P9 IUSS 1 135.0°C (275.0°F) 3 min 1 min Full instrument packs 1 1 1 Vac Full fabric packs 1 1 1 P10 IUSS 2 Grav 135.0°C (270.0°F) 10 min 30 sec Full instrument packs 1 1 1 Full fabric packs 1 1 1 P11 IUSS 3 Grav 132.2°C (270.0°F) 4 min 1 min Full instrument packs 1 1 1 Full fabric packs 1 1 1 P12 Vented Bottles 121.1°C (250.0°F) 45 min 3kPa/min (0.44 psi/min) (Note 2) Each container
1000 mL (34 fl oz) or smaller (Note 3) 3 3 3 NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact. TABLE NOTES:

1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAMI Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8). For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to steam sterilization and sterility assurance in health care facilities.

2. Cooldown rate

3. Use vented or open containers only.

See the full product information製品情報をすべて見るには