Standard Sterilizable Tray
Implant Direct Sybron Manufacturing, LLC
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The Standard Sterilizable Tray is designed to hold various dental surgical and prosthetic instruments in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap (maximum thickness KC300) and sterilized in an FDA cleared sterilizer for one of the following cycles:
(1) Prevacuum Steam – At 132°C for 4 minutes with a 20 minutes dry time. (2) Gravity Steam – At 132°C for 15 minutes with a 30 minutes dry time.
- The tray is intended for sterilization of non-porous loads.
- Do not stack trays during sterilization.
- The tested Tray represents the worst case validated load of 667.52 grams.
- Implant Direct Sybron Manufacturing LLC does not make any lumen claims for the Standard Sterilizable Tray.