TrueView 100 Pro
Compai Healthcare (Shenzhen) Co.,Ltd
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
A cabinet X -ray system used to provide digital X -ray images of surgical and core biopsy specimens from various anatomical regions in order to allow rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification in the same room as the procedure or nearby improves workflow, thus reducing the time the patient needs to be under examination.
Standards: IEC 61010- 1:2010 – Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements IEC 61010-2-091:2012 – Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 2- 091: Particular Requirements for Cabinet X-ray Systems IEC6101-2-101:2018 – Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2 -101: Particular requirements for in vitro diagnostic (IVD) medical equipment EN 61326-1:2013 – Electrical Equipment for Measurement, Control, and Laboratory Use – EMC Requirements, General Requirements EN 61326-2-6:2013 – Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-6: Particular requirements for in vitro diagnostic (IVD) medical equipment EN 55011 (Class A) – Industrial, Scientific and Medical (ISM) Radio-Frequency Equipment – Electromagnetic Disturbance Characteristics – Limits and Methods of Measurement CompAI Healthcare (Shenzhen) Co., Ltd 8B,Huangting Building, No.355, Fuhua Road,Futian Street,Futian District,Shenzhen,P.R.China 5-3 ISO 14971: 2019 – Medical devices – Application of Risk Management to Medical Devices IEC 62366- 1:2015 – Medical devices - Part 1: Application of usability engineering to medical devices IEC 62304:2006/AMD1:2015 – Medical device sof tware - Software life cycle processes FDA Guidance Documents: “Applying Human Factors and Usability Engineering to Medical Devices,” issued on February 3, 2016 “Guidance for the Submission of 510(k)s for Solid State X -ray Imaging Devices,” issued on September 1, 2016 “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices,” issued on May 11, 2005 “Cybersecurity for Networked Medical Devices Containing Off -the-Shelf (OTS) Software,” issued on January 14, 2005 “Off-the-Shelf Software Use in Medical Devices,” issued on September 9, 2019 “Compliance Guide for Cabinet X-Ray Systems” issued on September 19, 2007 Format for Traditional and Abbreviated 510(k)s Guidance for Industry and Food and Drug Administration Staff
Comparison with Predicate Device: The TrueView 100 Pro and its predicate device, the Trident® HD Specimen Radiography System, have the same intended use, general configuration, principles of operation, and operating parameters.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。