DECISION DATEJanuary 12, 2021PRODUCT CODEFXX

Indications for use

The Surgical Face Mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials. Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)) is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials. Same Mask style Flat-pleated, 3 layers Flat-pleated, 3 layers Same Design feature Ear loop/Tie-on Ear loop/Tie-on Same Color Blue Blue Same Dimension

1. 14.5cm×9cm: Nose Clip: 85mm×2.9mm Ear Loop: 180mm×3mm Ties: 910mm×10mm

2. 17.5cm×9.5cm: Nose Clip: 100mm×2.9mm Ear Loop: 180mm×3mm Ties: 910mm×10mm 17.5cm×9.5cm Different ASTM F2100 Level Level 2 Level 2 Same Sterility Sterile/Non-sterile Non-sterile Different Use Single Use, Disposable Si ngle Use, Disposable Same Fluid resistance Pass at 120mmHg Pass at 120mmHg Same Particulate filtration efficiency Average 98.98% Average 99.74% at 0.1μm Similar Bacterial filtration efficiency Average 98.92% Average 99.4% Similar Differential pressure Average 4.4 mmH 2O/cm2 Average 2.7mmH 2O/cm2 S i m i l a r Flammability Class 1 Class 1 Same Label/Labeling Complied with 21 CFR part 801 Complied with 21 CFR part 801 Same Patient Contacting Material Outer facing layer Spun-bonded nonwoven polypropylene Spun-bond polypropylene Different Middle layer Melt-blown nonwoven polypropylene Melt blown polyp ropylene filter Inner facing layer Spun-bonded nonwoven polypropylene Spun-bond polypropylene nose clip Medical polypropylene and Q235 Malleable aluminum wire Ear loops Nylon and spandex Polyester Tie-on Spun-bonded nonwoven polypropylene Spun-bond polypropylene Biocompatibility ISO 10993-5 and ISO 10993-10; Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing, and non-irritating. ISO 10993-5 and ISO 10993-10; Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing, and non-irritating. Same

Different– Dimension The dimension for proposed device is different from predicate d evice. The proposed device provides more dimension options, which can allow users to choose the appropriate dimension of masks according to their needs. In addition, the different size does not affect the indication for use.

Different- Sterility The proposed device can be provided in sterile and non-sterile. The proposed non-sterile mask is same as the predicate device, while the sterile mask is not covered by the predicate device. However, the sterilization process were validated per ISO 11135 and the resu lt demonstrated that the sterilization process is effective. In addition, the performance test and bio compatibility test were also conducted on the sterilized mask and the test result demonstrated that the d evice can meet the requirements of level 2 mask and does not raise any adverse effects after sterilization.

Similar –Particulate filtration efficiency The test result for particulate filtration efficiency for the p roposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask.

Similar - Bacterial filtration efficiency The test result for bacterial filtration efficiency for the pro posed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask.

Similar - Differential pressure The test result for different pressure for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask.

Different- Patient Contacting Material The patient contacting material for the proposed device is diff erent from predicate device. However, biocompatibility test has been performed on the proposed device and the results does not show any adverse effect.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。