RARA CRAFT DATABASE for FDA Medical Devices
510(k)K203784

FDA DEVICE FACTS

PreXion3D Explorer PRO

Prexion Corporation

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEFebruary 23, 2021PRODUCT CODEOAS

Indications for use

PreXion3D Explorer PRO is intended to produce two dimensional digital panoramic and cephalometric images, and three-dimensional digital x-ray images of the maxillofacial, and ENT (Ear, Nose and Throat) region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.

This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year-old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing PreXion3D Explorer EX is intended to produce two dimensional digital panoramic and cephalometric images, and three-dimensional digital x-ray images of the maxillofacial, and ENT (Ear, Nose and Throat) region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.

This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year-old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing PreXion3D Explorer is intended to produce two‐dimensional digital x‐ray images including panoramic and cephalometric image, and three‐dimensional digital x‐ray images of the dental, oral, maxillofacial region,ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment. Similar PreXion Corporation Traditional 510(k) Premarket Submission PreXion3D Explorer PRO Section 5 radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients. radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients. Patient/User Characteristics Target Population Children aged 6 (except infants) to elderly Same Same Same Anatomical Site The dental, oral, maxillofacial region ENT (Ear, Nose and Throat) and neck region Same Same Same Users Health care professionals Same Same Same Technological Characteristics and Performance Patient Contact Material CHIN REST: polycarbonate Forehead Holder: silicone rubber HANDLE GRIP: silicone rubber Same CHIN REST: polycarbonate HEADREST: carbon BELT: polyvinyl chloride Biocompatibili ty test results support that the material is safe for the intended use. Sterility Non-sterile Same Same Same X-ray Generati on Device Tube Voltage 90-110KV 90-110KV 90-110KV Same Pulse Exposure function Yes Yes Yes Same Tube Current 1-5.3mA 1-5mA 1-5mA Same Focal Spot Size 0.3mm x 0.3mm Same Same Same X-ray Image Capturin g Device Detector FPD (TFT) Same Same Same Pixel Size 248 µm x248µm (With binning) 248 µm x248µm (With binning) Same Same as secondary

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 24, 2026