Steam-Dot Process Indicator
Propper Manufacturing Co., Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The Steam-Dot process indicator for steam sterilization is designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. The indicator dots turn from white to dark brown/black when exposed to steam sterilization conditions, thus providing an indication of processed items.
The Steam-Dot process indicator can be used in the following steam sterilization cycles: Gravity: 121oC/250oF - 30 min Pre-vacuum: 132oC/270oF -3min Pre-vacuum: 132oC/270oF -4min Pre-vacuum: 134oC/273oF -3 min Pre-vacuum: 134oC/273oF -4 min Pre-vacuum: 135oC/275oF -3 min The Process Indicator Tape for Steam Sterilization is indicated for use in holding sterilization packs together and can be used in gravity sterilizers operating at 121°C for 30 minutes or pre-vacuum sterilizers operating at 132°C for 4 minutes and 135°C for 3 minutes. The indicator stripes turn dark brown/black when exposed to steam sterilization conditions, thus providing an indication of processed items. Similar
Steam-Dot Process Indicator can be used in additional cycles: 132oC-3 min pre-vacuum, 134oC-3 min pre-vacuum 134oC-4 min pre-vacuum.
The predicate device is a process indicator tape that changes color in the presence of the sterilant at Gravity: 121oC for 30 minutes and Pre-Vacuum: 132oC for 4 minutes and 135oC for 3 minutes. Summary of Non-Clinical Testing
Test Purpose Acceptance Criteria Result ANSI/AAMI/ ISO 11140- 1:2014 testing for Type 1 indicator. To demonstrate conformance of Steam-Dot indicator to the requirements specified in ISO 11140-1:2014 for process indicators. 121oC-10 min: dark brown or black color 121oC-2 min: no color change or color markedly different compared to dark brown or black 134oC-2 min: dark brown or black color 134oC-0.3 min: no color change or color markedly different compared to dark brown or black 140oC-30 min Dry heat: no color change Passed FDA Guidance for industry for chemical indicators. Steam process indicator performance test. To demonstrate conformance of Steam-Dot indicator to the requirements specified in the FDA Guidance for industry for process indicators. 121oC-10 min: dark brown or black color 121oC-2 min: no color change or color markedly different compared to dark brown or black 132-135oC-2 min: dark brown or black color 132-135oC-20sec: no color change or color markedly different compared to dark brown or black 140oC-30 min Dry heat: no color change Passed Testing in hospital type sterilizers in gravity and pre-vacuum 510k cleared cycles. To demonstrate that Steam- Dot Process indicator achieves specified end color in typical cycles in hospital sterilizers. Color change from white to dark brown or black Passed Biocompatibilit y study and ink transfer test To demonstrate that the indicator does not create biocompatibility issues to health care professionals and patients. Individual components should not create biocompatibility issues. Testing according to ISO 11140-1:2014. Requirement: 6.2.2. No ink transfer should be observed on unprocessed and steam processed indicators. Passed End point stability and shelf- life study To confirm that Steam-Dot process indicator has acceptable stability after processing when achieved and not achieved end point color (“Pass” and “Fail” conditions).
To demonstrate that Steam- Dot indicator meets the performance parameters when tested using real-time shelf-life exposure method. Steam-Dot indicators processed in Pass and Fail cycles at various time points after production and at the end of shelf life should demonstrate stable color for 110 days.
Meet specifications after real-time 48 months shelf-life exposure. Passed Adhesive test – healthcare applications The purpose of the test is to evaluate if the adhesive is suitable for its function and does not deteriorate during sterilization process. The test is considered a pass if at least 95% of the indicators at different stages of shelf life remain on surfaces of sterilization packaging materials before and after steam processing. Passed
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。