SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate

STERIS Corporation

Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。

DECISION DATEMay 12, 2021PRODUCT CODEMED

Indications for use

The SYSTEM lE Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities. The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use. The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices. * The extensive treatment of EPA potable water consists of: 1. Pre-filtration through two pre-filters:

• Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.

• Pre-filter Bis a surface filter that removes particles/contaminants> 0.1 micron. 2. UV Irradiation:

• During transit through the UV water treatment chamber, a UV dose sufficient to achieve a> or equal to 6-log reduction ofMS2 virus is delivered to the water. 3. 0.1 micron filtration:

• The water prepared by pre-filtration and UV irradiation is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa> 0.1 micron.

The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities. The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution(> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water. The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 23, 2026