FDA DEVICE FACTS
Auxein Nailing System
Auxein Medical Private Limited
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
AUXEIN Brand of Expert Tibia Nailing System AUXEIN Brand of Expert Tibia Nailing System- consists of variety of Bone nails and bone screws. Expert Tibia Nail System is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions. AUXEIN Brand of Expert Femoral Nailing System AUXEIN Brand of Expert Femoral Nailing System -is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions. AUXEIN Brand of Gamma Nailing System AUXEIN Brand of Gamma Nailing System -The Gamma Nail with Anti -Rotation Screw is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections. AUXEIN Brand of AJAX Advance Nailing System, AUXEIN Brand of AJAX Advance Nailing System, end cap and screws is intended to treat stable and unstable proximal femoral fracture including Pertrochanteric fractures, Intertrochanteric fractures and High sub trochanteric fractures and combination of these fractures. The AUXEIN MEDICAL's Elastic Intramedullary Nail System The AUXEIN MEDICAL's Elastic Intramedullary Nail System is indicated for fixation of diaphyseal fractures where the canal is narrow or flexibility of the implant is important. This includes upper extremity fractures in all patients and lower extremity fractures in children above the age of 6 years and any patient below 65 kg. This system is also intended to treat metaphyseal and epiphyseal fractures, such as radial neck fractures. AUXEIN Brand of Humeral Nailing System- AUXEIN Brand of Humeral Nailing System is indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures, proximal 510K Number: K210792
humeral fractures with diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures.. AUXEIN Brand of Retrograde Nailing System AUXEIN Brand of Retrograde Nailing System This nails are indicated for simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated.” (A.6). Summary of Technological Characteristics as compared to the predicate devices: Substantial equivalence including comparison with predicate devices. A comparison between the Auxein Nailing System and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria. Following is the summary of parameters in which the comparison has been verified:
S.No. Characteristics Predicate Device Versus New Device (Auxein Brand) Remarks
1. Indications for use Identical Indications for use. Equivalent
2. Material Same material used in New Device and Predicate device. Equivalent
3. Performance Standards Same performance standards used in both New Device as well as predicate device. Equivalent
4. Sterilization (Gamma) Earlier the predicate device was supplied Non-sterile but the now device will be supplied Sterile. Equivalent
5. Dimensional Verification Same dimensions are found in both New Devices as well as Predicate device. Equivalent
(C.1). Discussion on the non-clinical testing performed Following are the applicable product standards considered for non-clinical standards: A: Material Standards B: Performance Standards
A: Material Standards: The material standards are the essential part to be complied with first, as it is the basis of 510K Number: K210792
manufacturing metallic surgical implants. We have complied with the following material standards. 1. ASTM F 136: Standard specification for wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra low interstitial) Alloy for surgical implant applications. 2. ASTM F 899-12: Standard Specification for Wrought Stainless Steels for surgical instruments. We have verified the purchased material compliance to these standards and copies of the relevant test results are attached herewith.
B: Performance Standards: The device's performance of Auxein Nailing System has been demonstrated against following applicable standards.
● ASTM F 543-17
● ASTM F1264
● ASTM F384
● Custom Fatigue Testing
(B.1). Discussion on the clinical evaluation referenced and relied upon: Auxein Nailing System is of similar design and pattern as well as similar intended use.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。