RARA CRAFT DATABASE for FDA Medical Devices
510(k)K210836

FDA DEVICE FACTS

InstruSafe Instrument Protection System

Summitt Medical, LLC

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATENovember 30, 2021PRODUCT CODEKCT

Indications for use

InstruSafe® Instrument Protection System cassettes(trays) are used to organize and protect other medical devices that are sterilized by a healthcare provider. lnstruSafe Instrument Protection System cassettes are intended to allow sterilization of the encl osed medical devices during STERIZONE VP4 sterilization Cycle 1. The lnstruSafe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The InstruSafe Instrument Protection Sys tem cassettes are not intended on their own to maintain sterility. Summit Medical has validated the use of the InstruSafe Instrument Protection System cassettes in Cycle 1 of the STERIZONE VP4 sterilizer through demonstrations of sterilization efficacy using representative samples of medical devices including general instruments, rigid and semi-rigid instruments with lumens, and flexible endoscopes. The validation provided information that has been used to establish design limits that are applied across the range of InstruSafe Instrument Protection System cassettes to ensure that all models fall within the validated limits for cassette ventilation, internal shelving, and weight:

• Minimum ventilation-to-volume ratio, general instruments: 0.073

• Minimum ventilation-to-volume ratio, flexible endoscopes: 0.263

• Maximum number of internal shelves: Two (2)

• Maximum cassette weight (including contents): 25 lbs/cassette Refer to Table for the validated loads using InstruSafe Instrument Protection System cassettes. Ref er to the sterilizer manufacturer’s instructions for use to ensure that loads do not exceed the claimed limits for the sterilizer. Table – Description of InstruSafe Instrument Protect System Loads. Representative of STERIZONE® VP4 Sterilizer Validation Load No. (K190260) Summit Medical Validation Load Description Total Load Weight (excludes the 25-lbs. loading rack) Load 4a: Consisted of the load limit for rigid and semi-rigid channeled instruments:

• Three (3) double channel semi-rigid endoscopes (ureteroscope – 0.7 mm × 500 mm and 1.1 mm × 500 mm) were packaged, one (1) per cassette, in three (3) IN-7323-R cassettes.

• Additional rigid channel instruments were added to reach a total of 15 channels. Each cassette included appropriate silicone instrument holders and was wrapped. Load 4b: Consisted of the load limit for rigid and semi-rigid channeled instruments:

• Three (3) double channel semi-rigid endoscopes (ureteroscope – 0.7 mm × 400 mm and 1.1 mm × 400 mm) were packaged, one (1) per cassette, in three (3) IN-0007-TF cassettes.

• Additional rigid channeled instruments were added to reach a total of 15 channels. Each cassette included appropriate silicone instrument holders and was placed in an Aesculap JM440 rigid container. Load 4a = 5 lbs/cassette, 11 lbs/cassette, Representative of STERIZONE® VP4 Sterilizer Validation Load No. (K190260) Summit Medical Validation Load Description Total Load Weight (excludes the 25-lbs. loading rack) Summit Medical Validation Load 7a: Consisted of the load weight limit for general medical instruments representing the following geometries:

• Box-lock hinge

• Pivot hinge

• Luer-lock General medical instruments were spread across three (3) IN-7323-R cassettes with appropriate silicone holders, each weighing 25 lbs. The cassettes were wrapped. Summit Medical Validation Load 7b: Consisted of the load weight limit for general medical instruments representing the following geometries:

• Box-lock hinge

• Pivot hinge

• Luer-lock General medical instruments were spread across three (3) IN-0006-TF cassettes with appropriate silicone holders. Cassettes were placed in Aesculap JM440 containers, each weighing 25 lbs. Load 7a = 25 lbs/cassette 25 lbs/cassette Summit Medical Validation Load 8a: Consisted of the load limit of five (5) total flexible endoscope channels in wrapped cassettes:

• Two (2) double-channel flexible endoscopes (ureteroscopes

– 1 mm x 850 mm and 1 mm x 989 mm) were packaged, one (1) per cassette, in IN-0003-R cassettes.

• One (1) single-channel flexible endoscope (ureteroscope – 1 mm x 850 mm) was packaged in an IN-0004-R cassette. Cassettes included appropriate silicone holders and were wrapped. Summit Medical Validation Load 8b: Consisted of the load limit of four (4) total flexible endoscope channels in cassettes packaged in rigid containers:

• Two (2) double channel flexible endoscopes (ureteroscope – 1 mm x 850 mm and 1 mm x 989 mm) were packaged, one (1) per cassette, in IN-0003-TF cassettes with appropriate silicone holders. Each cassette was placed inside an Aesculap JM442 container. Load 8a = 4.3 lbs/cassette, 8.3 lbs/cassette,

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 23, 2026