FDA DEVICE FACTS
Plus Medical Chemotherapy Gown
Foshan Nanhai Plus Medical Co, Ltd.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
Plus Medical Chemotherapy Gowns are intended to protect healthcare personnel from exposure to chemotherapy drugs during preparation, handling and administration. The gowns are a closed back design, single use, disposable medical device provided sterile and non-sterile. Non-sterile gowns are to be sold in bulk to re-packager/re-labeler establishments for ethylene oxide (EtO) sterilization according to ISO 11135-1 prior to marketing to the end users and Sterile surgical gowns are to be sold directly to users after EtO sterilization validation to ISO 11135-1.
Plus Medical Chemotherapy Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012. Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities.
Plus Medical Chemotherapy Gown has been evaluated for resistance to permeation of various chemotherapy drugs per ASTM F739-12 Standard Test Method for Permeation of Liquids and Gases through Protective Clothing Materials under Conditions of Continuous Contact. Testing data is provided per test procedure ASTM F739-12, the following 12 chemotherapy drugs has no breakthrough at the Standardized Breakthrough Rate of 0.1ug/cm2/minute, up to 480 minutes.
Test Chemotherapy Drugs Average Standardized Breakthrough Time
Carmustine (BCNU) [3.3mg/ml] >480mins
Cisplatin [1.0mg/ml] >480mins
Cyclophosphamide (Cytoxan) [20.0mg/ml] >480mins
Dacarbazine [10.0mg/ml] >480mins
Doxorubicin HCI [2.0mg/ml] >480mins
Etoposide [20.0mg/ml] >480mins
Fluorouracil [50mg/ml] >480mins
Methotrexate [25mg/ml] >480mins
Mitomycin C [0.5mg/ml] >480mins
Paclitaxel [6.0mg/ml] >480mins
Thio Tepa [10mg/ml] >480mins
Vincristine Sulfate [1.0mg/ml] >480mins
*No permeation was detected at either the minimum detectable permeation or 0.lug/cm2/minute.
When chemotherapy drugs are present, gown selection should be based on the specific type(s) of chemicals used. Users are recommended to review drug labeling or material safety data sheets for the chemicals being used to determine an adequate level of protection.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。