Indications for use
SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging using a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion. Note: The intended use of SenoBright HD cleared in K172404 is not changed. Indications for Use Refer to Intended use. Technology: The update of the existing recombination algorithm to New Image Recombination Algorithm (NIRA) for SenoBright HD is completely independent from all other parts of the digital mammographic equipment Senographe Pristina. It therefore does not affect the dual-energy image acquisition.
The changes introduced with NIRA are specific to the recombination algorithm and can affect exclusively the recombined image. Thus, it also does not affect the acquisition, processing and display of the low-energy image when used as a regular FFDM image.
NIRA is an evolution of the SenoBright HD Recombination algorithm bringing the following improvements:
- Use of a local estimation of breast thickness in the image recombination to account for the non-uniformity of the breast thickness.
- Compensation for potential patient movement between the 2 CESM acquisitions (Low Energy and High Energy). Substantial Equivalence /