VSP® Orthopedics System

3D Systems

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATENovember 10, 2021PRODUCT CODEPBF

Indications for use

The VSP® Orthopedics System is intended to be used as a surgical instrument to assist in preoperative planning and/or in guiding the marking of bone and/or in guiding surgical instruments in non‐acute, non‐joint replacing osteotomies for adult patients in the distal femur, tibia, and non‐sacrum pelvis.

7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The intended use and technological characteristics of the subject device are either identical or substantially equivalent to the predicate device (VSP® Orthopedics System), differing only in the expansion of materials included within the VSP® Orthopedics System for anatomic models.  The potential impact on substantial equivalence of each technological difference was addressed by risk analysis and verification and validation testing. Similarities to Predicate The inputs to the subject device are identical to those of the predicate device. System outputs of the subject and predicate device comprise both digital and physical outputs which can include patient‐specific anatomical models, and guides for the purpose of assisting the surgeon during maxillofacial surgeries.  The subject device and the predicate device are both intended to be generated by trained personnel, in a non‐medical manufacturing or office environment, with active support from the surgeon.  The subject device uses the same technologies as the predicate device.  These include image transfer and manipulation via software that is subsequently used for 3D printing of anatomical models and guides for surgical planning. The subject and predicate device use the same software components for the digital workflow.  There has been no change to the sterilization process for the surgical guides produced by the subject device. The subject device serves the same patient population as the predicate device. Differences to Predicate One new material is presented in the modified system and makes up the changes presented within this Traditional 510(k) submission.  The material, SLA printing resin Accura ClearVue, is being added to the system for anatomical models.

8. SUMMARY OF PERFORMANCE TESTING The testing outlined below was intended to show that the output of the design and development process demonstrated compliance with the device specifications. Non‐clinical testing was conducted to prove the subject device performs in accordance with its intended use and is substantially equivalent to the listed predicate device. K211244 Page 3 of 3 The following testing was conducted for the VSP® Orthopedics System:  Operational Qualification A process operational qualification was performed to assess the manufacturing process over the range of allowable process parameters. All test method acceptance criteria were met.  Performance Qualification A process performance qualification was conducted to assess the manufacturing process repeatability at nominal parameters. All test method acceptance criteria were met.  Cleaning Validation A cleaning validation was performed in accordance with AAMI TIR 30. Following soiling and cleaning of the subject device, bioburden, protein levels, and hemoglobin levels were exampled. All test method acceptance criteria were met.  Sterilization Validation Steam sterilization validations were performed for the dynamic‐air‐removal cycle in accordance with ISO 17665‐1:2006 to a sterility assurance level (SAL) of 10‐6 using the biological indicator (BI) overkill method. All test method acceptance criteria were met.  Biocompatibility Validation Biocompatibility endpoints were evaluated in accordance with ISO 10993‐1. Sensitization, irritation, and acute systemic toxicity testing conducted on the subject device manufactured from Accura ClearVue are within the pre‐defined acceptance criteria. The results of the testing adequately address biocompatibility for the output device and it’s intended use.  Shelf Life Validation A shelf life validation was performed to ensure Accura ClearVue anatomical model outputs maintain functionality following a shelf life of 3 months.  Clinical Performance Data Clinical testing was not necessary for the determination of substantial equivalence. All process and verification and validation testing, which were conducted as a result of risk analyses and design impact assessments, showed conformity to pre‐established specifications and acceptance criteria. The acceptance criteria was established in support of device performance, and testing demonstrated substantial equivalence of the system to the predicate device.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン