FDA DEVICE FACTS
LED PHOTIC System
Micromed S.P.A.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
LED PHOTIC System is an accessory indicated for photic activation of brain during the electroencephalographic (EEG) studies. It always has to be used in combination with an EEG system, in order to evaluate particular anomalies of the brain activity due to the intermittent photic stimulation. The product must be used only by qualified personnel (physicians or technicians of neurophysiopathology) for the execution of EEG exams in a professional environment. Comparison with Predicate Device:
Device & Predicate Device(s): K211974 K101691 Comments General Device Characteristics
Intended Use LED PHOTIC System is an The Lifelines Proposed IFU accessory indicated for photic activation of brain during the Photic Stimulator is contains additional electroencephalographic (EEG) studies. It always has to indicated for photic activation information, but IFU intent be used in combination with an EEG system, in order to of the EEG during an remains the same. evaluate particular anomalies of the brain activity due to the intermittent photic stimulation. The product must be used only by qualified personnel (physicians or technicians of neurophysiopathology) for the execution of EEG exams in a professional environment. EEG study and in the generation of visual evoked potentials.
Light output Max 1.9 kLux at 10 cm/Max 1.1 kLux at 20 cm/Max 0.7 kLux at 30 cm 13 kLux at 1 ft. Predicate device has lower Lux at similar distance. Flash rate 0-60Hz 1-60Hz Similar Power supply 120Vpower (wall outlet) access required. Can be powered via USB on host PC or power supply. Similar Software Computer controlled (not included) Computer controlled (software included) Predicate device has software capabilities.
Performance Testing: The LED PHOTIC System was tested and found compliant to the following safety standards:
• IEC 60601-1:2005 + A1:2012: Medical Electrical safety
• IEC 60601-1-6:2010 + A1:2013: Usability
• IEC 60601-2:2014: EMC Compliance
• IEC 62471:2006: Photobiological Safety