RARA CRAFT DATABASE for FDA Medical Devices
510(k)K212200

FDA DEVICE FACTS

STERLINKTM FPS-15s Plus Sterilizer with STERLOADTM Cassette

Plasmapp Co,., Ltd.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEOctober 22, 2021PRODUCT CODEMLR

Indications for use

The STERLINK ™ FPS-15s Plus sterilizer with STERLOAD ™ cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre -programmed sterilization lumen cycle operates at low pressure and low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

The STERLINK ™ FPS-15s Plus can sterilize*:

• Instruments with diffusion -restricted spaces such as the hinged portion of forceps and scissors

• Medical devices with a single stainless -steel lumen with: o An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

*The validation testin g for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 3.97 lbs. The PSD -85 is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities.

The two pre -programmed sterilization cycles, the Lumen and the Non- Lumen Cycles, operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

The PSD-85 Lumen Cycle can sterilize*:

• Instruments with diffusion -restricted spaces such as the hinged portion of forceps and scissors

• Medical devices with a single stainless steel lumen with: o An inside diameter of 1 mm or larger and a length of 60 mm or shorter o An inside diameter of 2 mm or larger and a length of 250 mm or shorter o An inside diameter of 3 mm or larger and a length of 350 mm or shorter

*The validation testing for all lumen sizes was conducted using a maximum of ten (10) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by thi s testing. The validation studies were performed using a validation load consisting of one instrument tray with a total weight of 10.4 lbs.

The PSD -85 Non- Lumen Cycle can sterilize**:

• Non -lumen instruments including non-lumen instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.

**The validation studies were conducted using a validation load consisting of one instrument tray with a total weight of 25.3 lbs. Physical Characteristic Self-contained, stand-alone device Same Design and Construction Welded frame onto which is mounted the sterilization chamber along with a variety of instruments and components, controls, piping, and vacuum pump, all of which is housed in covered frame Same STERLINK™ FPS-15s Plus Sterilizer with STERLOAD™ Cassette Subject Device Predicate Device Plasmapp Co., Ltd. Sterilucent, Inc. Chamber Volume 14 L 85 L Weight 147 lbs (67 kg) 440 lbs (200 kg) Max Power 1000 W 1650 W Control System Embedded Linux Atmel AT91SAM7 Microprocessor Internal process monitor Temperature Chamber and vaporizer thermocouple Chamber and vaporizer thermistors Pressure Chamber pressure transducers Same Sterilant Concentration None Real-time hydrogen peroxide vapor monitor Operational Principle Sterilization of the intended devices by exposure under controlled conditions of pressure, temperature, and time Same Operational Parameters Low pressure (vacuum; sub -atmospheric down to 3 Torr) and temperature (60°C) Low pressure (vacuum; sub -atmospheric down to 0.4 Torr) and temperature (55°C) Pre-processing Requirements Cleaned, rinsed, and dried devices Same Devices Reusable metal and non -metal medical devices that are used in healthcare facilities, including those that are sensitive to heat and moisture Same Sterilization Cycles One (1) pre-programmed; approximately 36 minutes Two (2) pre-programmed; Lumen (approximately 72 minutes) and Non-Lumen (approximately 40 minutes) Sterilant Model Name STERLOAD™ Sterilant Disc Type Cassette type (unit dose) Disk type Sterilant 59% aqueous solution of hydrogen peroxide Same Monitoring accessories Biological Indicator Self-contained biological indicator, Geobacillus stearothermophilus Same Process Challenge Device / Routine Test Pack Self-contained biological indicator, Geobacillus stearothermophilus Same Chemical Indicator Terragene® CI Strips and Tapes Sterilucent CI Strips, Labels and Tape Miscellaneous (Sterilization wrap) Load Packaging Tyvek®/HDPE pouches Tyvek®/Mylar pouches, trays wrapped with polypropylene sterilization wrap, rigid containers STERLINK™ FPS-15s Plus Sterilizer with STERLOAD™ Cassette Technological differences between the STERLINK ™ FPS-15s Plus sterilizer and the predicate device are minimal and include:

• The STERLINK™ FPS-15s Plus sterilizer has a smaller sterilization chamber volume, is lighter, and requires less maximum power compared to the predicate device.

• The STERLINK™ FPS-15s Plus sterilizer has one pre-programmed sterilization lumen cycle, versus two cycles for the predicate device.

• The STERLINK™ FPS-15s Plus sterilizer utilizes unit dose sterilant cassettes versus sterilant disks for the predicate device.

• The STERLINK™ FPS-15s Plus sterilizer utilizes a fixed dose of sterilant, versus a variable dose of sterilant based on real time measured and monitored hydrogen peroxide vapor concentration for the predicate device.

The STERLINK™ FPS-15s Plus sterilizer performance and safety characteristics have been shown to provide a level of safety and efficacy at least equivalent to that of the predicate device.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 23, 2026