RARA CRAFT DATABASE for FDA Medical Devices
510(k)K212424

FDA DEVICE FACTS

AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam Sterilizers

STERIS Corporation

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEOctober 28, 2021PRODUCT CODEFLE

Indications for use

The Amsco 400 Small Steam Sterilizer models 16, 16C, 16CS, 16S, 20, 20C, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

• Prevacuum (Models 16, 16C and 20, 20C) – are equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

• Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20S) – are equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The Amsco 400 Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-1): Table 4-1. Amsco 400 Small Prevacuum Steam Sterilizer Models 16, 16, 20, 20C factory-validated sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load Prevac 270°F (132°C) 4 minutes 20 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. Prevac 270°F (132°C) 4 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities. Prevac 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack Prevac 275°F (135°C) 3 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Prevac (Immediate Use) 270°F (132°C) 4 minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity 250°F (121°C) minutes 15 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. Gravity 270°F (132°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity 250°F (121°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity 270°F (132°C) minutes 15 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. Gravity (Immediate Use) 270°F (132°C) 3 minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. *The liquid cycle is for non-patient contact use only. The Amsco 400 Small Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-2): Table 4-2. Amsco 400 Small Steam Flush Pressure-Pulse (SFPP)Sterilizer Models 16S, 16CS, 20S, 20CS factory-validated sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load SFPP 270°F (132°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities. SFPP 270°F (132°C) minutes 20 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. SFPP 275°F (135°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. SFPP 270°F (132°C) minutes 5 minutes Single Fabric Pack Prevac 270°F (132°C) minutes 20 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. Prevac 270°F (132°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities. Prevac 270°F (132°C) minutes 5 minutes Single Fabric Pack Prevac 275°F (135°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Prevac (Immediate Use) 270°F (132°C) minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity 250°F (121°C) minutes 15 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. Gravity 270°F (132°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity 250°F (121°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity 270°F (132°C) minutes 15 minutes Fabric Packs. Refer to Table 4-3 for recommended quantities. Gravity (Immediate Use) 270°F (132°C) minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Liquid* 250°F (121°C) minutes N/A Validated with 3 1000ml bottles. Refer to Table 4-4 for recommended full load quantities. DART Warm-Up 270°F (132°C) 3 minutes 1 minute N/A DART 270°F (132°C) 3 ½ minutes 1 minute Bowie-Dick Test Pack, DART Test Pack Leak Test N/A N/A N/A N/A Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load Gravity (Immediate Use) 270°F (132°C) minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Gravity (Immediate Use) 270°F (132°C) minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities. Liquid* 250°F (121°C) minutes N/A Validated with 3 1000ml bottles. Refer to Table 4-4 for recommended full load quantities. DART Warm-Up 270°F (132°C) minutes 1 minute N/A DART 270°F (132°C) 3 ½ minutes 1 minute Bowie-Dick Test Pack, DART Test Pack Leak Test N/A N/A N/A N/A *The liquid cycle is for non-patient contact use only. The following table lists STERIS’s recommended loads by sterilizer size: Table 4-3 Amsco 400 Small Steam Sterilizer recommended loads per sterilizer size Sterilizer Size Wrapped Instrument Trays Fabric Packs 16” x 16” x 26” 2 4 20” x 20” x 38” 3 6 The following table is a guideline for liquid cycle processing: Table 4-4 Amsco 400 Small Steam Sterilizer Liquid* Cycle Guideline Sterilizer Size Number of Containers for full load Number of Containers for validation Volume of Liquid In One Container Minimum Recommended Sterilize Time at 250ºF (121ºC) 16” x 16” x 26” 15 3 1000 ml 45 minutes 20” x 20” x 38” 32 3 1000 ml 45 minutes *The liquid cycle is for non-patient contact use only. The Amsco 400 Small Steam Sterilizer is offered in the following chamber sizes:

• 16” x 16” x 26” (Models 16, 16C, 16CS, and 16S)

• 20” x 20” x 38” (Models 20, 20C, 20CS, and 20S) The AMSCO 400 Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

• Prevacuum – equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The AMSCO 400 Medium Prevacuum Steam Sterilizers are equipped with the following factory-programmed sterilization cycles and cycle values (Table 4.5): Table 4.5. AMSCO 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load Prevac 270°F (132°C) 4 minutes 20 minutes Fabric Packs. Refer to Table 4.6 for recommended quantities. Prevac 270°F (132°C) 4 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4.6 for recommended quantities. Prevac 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack Prevac 275°F (135°C) 3 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.6 for recommended quantities. Prevac (Immediate Use) 270°F (132°C) 4 minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table A-2 for recommended quantities. Gravity 250°F (121°C) 30 minutes 15 minutes Fabric Packs. Refer to Table 4.6 for recommended quantities. Gravity 270°F (132°C) 15 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.6 for recommended quantities. Gravity 250°F (121°C) 30 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.6 for recommended quantities. Gravity 270°F (132°C) 25 minutes 15 minutes Fabric Packs. Refer to Table 4.6 for recommended quantities. Gravity (Immediate Use) 270°F (132°C) 3 minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.6 for recommended quantities. Gravity (Immediate Use) 270°F (132°C) 10 minutes 1 minutes Unwrapped instrument tray with a single instrument or multiple unwrapped instrument trays, non-porous load, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.6 for recommended quantities. Liquid* 250°F (121°C) 45 minutes N/A Validated with 3 1000ml bottles. Refer to Table 4.7 for recommended full load quantities. DART Warm-Up 270°F (132°C) 3 minutes 1 minute N/A Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load DART 270°F (132°C) 3 ½ minutes 1 minute Bowie-Dick Test Pack, DART Test Pack Leak Test N/A N/A N/A N/A *The liquid cycle is for non-patient contact use only. The following table lists STERIS’ full loads by sterilizer size: Table 4.6 AMSCO 400 Medium Steam Sterilizer full load per sterilizer size Sterilizer Size Wrapped Instrument Trays Fabric Packs 26” x 26” x 39” 4 12 The following table is a guideline for liquid cycle processing: Table 4.7 AMSCO 400 Medium Steam Sterilizer Liquid Cycle Guideline Sterilizer Size Number of Containers for full load Number of Containers for validation Volume of Liquid In One Container Minimum Recommended Sterilize Time at 250ºF (121ºC) 26” x 26” x 39” 80 3 1000 ml 45 minutes *The liquid cycle is for non-patient contact use only. The AMSCO 400 Medium Steam Sterilizer is offered in the following chamber sizes:

• 26” x 26” x 39” (Models 39SL and 39CSL) The AMSCO 400 Medium Steam Sterilizers (Models 36H, 48H, 60H, 36SL, 48SL, 60SL, 36CH, 48CH, 60CH, 36CSL, 48CSL and 60CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:

• Prevacuum – equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles The Amsco 400 Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4.8): Table 4.8. Amsco 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Recommended Load Prevac 270°F (132°C) 4 minutes 20 minutes Fabric Packs. Refer to Table 4.9 for recommended quantities. Prevac 270°F (132°C) 4 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4.9 for recommended quantities. Prevac 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack Prevac 275°F (135°C) 3 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.9 for recommended quantities. Gravity 250°F (121°C) minutes 15 minutes Fabric Packs. Refer to Table 4.9 for recommended quantities. Gravity 270°F (132°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.9 for recommended quantities. Gravity 250°F (121°C) minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.9 for recommended quantities. Gravity 270°F (132°C) minutes 15 minutes Fabric Packs. Refer to Table 4.9 for recommended quantities. Liquid* 250°F (121°C) minutes N/A Validated with 3 1000ml bottles. Refer to Table 4.10 for recommended full load quantities. DART Warm-Up 270°F (132°C) 3 minutes 1 minute N/A DART 270°F (132°C) 3 ½ minutes 1 minute Bowie-Dick Test Pack, DART Test Pack Leak Test N/A N/A N/A N/A *The liquid cycle is for non-patient contact use only. The following table lists STERIS’s recommended loads by sterilizer size: Table 4.9 Amsco 400 Medium Steam Sterilizer recommended loads per sterilizer size Sterilizer Size Wrapped Instrument Trays Fabric Packs The following table is a guideline for liquid cycle processing: Table 4.10 AMSCO 400 Medium Steam Sterilizer Liquid Cycle Guideline Sterilizer Size Number of Containers for full load Number of Containers for validation Volume of Liquid In One Container Minimum Recommended Sterilize Time at 250ºF (121ºC) 26” x 37.5” x 36” 112 3 1000 ml 45 minutes 26” x 37.5” x 48” 154 3 1000 ml 45 minutes 26” x 37.5” x 60” 196 3 1000 ml 45 minutes *The liquid cycle is for non-patient contact use only. The AMSCO 400 Medium Steam Sterilizer is offered in the following chamber sizes:

• 26” x 37.5” x 36” (Models 36H, 36CH, 36SL, 36CSL)

• 26” x 37.5” x 48” (Models 48H, 48CH, 48SL, 48CSL)

• 26” x 37.5” x 60” (Models 60H, 48CH, 48SL, 48CSL)

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 23, 2026