FDA DEVICE FACTS
AceConnex Pre-Sutured Fascia
Allosource
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The AceConnex™ Pre-Sutured Fascia is intended for use as a component in soft tissue surgical procedures where constructs including those with allograft tissues are used for reconstruction, replacement, or augmentation of the hip labrum. Page 5-2 Technological Characteristics and Substantial Equivalence The AceConnex™ Pre-Sutured Fascia subject device is a pre-assembled surgical construct derived from allograft tissue and non-absorbable synthetic surgical suture. Donor tissue utilized in the subject AceConnex™ Pre-Sutured Fascia device meets eligibility requirements for Relevant Communicable Disease Agents or Diseases (RCDAD) via a medical director review of donor medical and social history and all applicable infectious disease screening tests, including human immunodeficiency virus type 1 (HIV -1 and HIV -1 NAT), human immunodeficiency virus type 2 (HIV -2), hepatitis B virus (H BV NAT, HBcAb , and HBsAg), hepatitis C virus (HCV and HCV -NAT), and syphilis (Rapid Plasma Reagin or Serologic Test for Syphilis). The tissue recovered from each individual donor is processed as an individual batch in a segregated manner to avoid cross -contamination with other donor tissue; at no time is tissue from multiple donors pooled. Donor tissue utilized in the AceConnex™ Pre- Sutured Fascia subject device meets the requirements of the American Association of Tissue Banks (AATB) and 21 CFR Parts 820 and 1271. The predicate device, Arthrex Suture Grafting Kit, is intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair. The AceConnex™ Pre -Sutured Fascia is substantially equivalent to this predicate device as the intended uses are the same. The subject device is a pre-assembled surgical construct , with an allograft tissue and suture component, and is intended to be used as a component in soft tissue surgical procedures where constructs are used for reconstruction, replacement, or augmentation of the hip labrum. Both devices contain nonabsorbable, synthetic, non-dyed suture. One difference between our device and this predicate, is that AceConnex Pre-Sutured Fascia is a pre-assembled sutured allograft , where the Arthrex Suture Grafting Kit is a device that is used in the assembly of sutured allografts in the operating room. This minor difference between AceConnex Pre - Sutured Fasc ia and th is predicate device does not raise any questions regardin g the safety and effectiveness and testing conducted has shown no apparent issue with the pre-assembly of the allograft in regards to the function and intended use of the device. Product Performance Testing Comprehensive product performance testing was conducted on the AceConnex ™ Pre-Sutured Fascia subject device to assess visual, physical, mechanical, and clinical properties . The subject device passed all product performance tests: device integrity was maintained during surgical preparation and application, device length was adjustable, the device conformed to the acetabular rim, the device was securable with knotted and knotless suture anchors, had a suture pullout strength greater than the estimated force on the labrum in a healthy hip during jogging, retains the necessary tensile strength after two years frozen storage, and has similar physical properties of surgeon fascia allografts. Bacterial Endotoxin Testing Bacterial endotoxin testing was conducted in accordance with the following standards:
• USP chapter <85> Bacterial Endotoxin Test
• ANSI/AAMI ST72:2011: Bacterial Endotoxins - Test methods, routine, monitoring, and alternatives to batch testing
• FDA Guidance for Industry: Pyrogen and Endotoxins Testing - Questions and Answers Bacterial endotoxin level s for the AceConnex™ Pre-Sutured Fascia subject device were tested using an adequate and acceptable representative endotoxin test sample for the device itself . Bacterial endotoxin levels were less than 20 EU/Device and met bacterial endotoxin testing requirements.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。