CoLink NeoFuse Plating System

In2bones USA, LLC

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEDecember 13, 2021PRODUCT CODEHRS

Indications for use

The CoLink® NeoFuse Plating System is a fusion plate indicated for anterior fixation of ankle arthrodesis and fractures, including the distal tibia and talus.

The addition of a compression screw through the tibiotalar joint (example CoLag 6.7mm screw) is required for ankle fusion procedure. The In2Bones USA LLC, CoLink® Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction of the small bones in the hand, wrist, foot and ankle in both pediatric and adult patients. The In2Bones NeoFuse Ankle Fusion plating system is indicated for anterior fixation of ankle arthrodesis and fractures, including the distal tibia, talus and calcaneus.

The addition of a compression screw through the tibiotalar joint (example IBS 6.5mm screw) is required. Product Code HRS, HWC HRS, HWC HRS, HWC Material Titanium Alloy (ASTM F136) Titanium Alloy (ASTM F136) Titanium Alloy (ASTM F136) Geometry and Dimensions

Available in 2 styles, 2 sizes each: Anterior and Anterolateral, 3 Hole and 5 Hole.

Screws: 3.5mm locking, non-locking or VAL screw or 4.5mm locking, non-locking or VAL screw Plate type: MTP, Lapidus, Y- Plate, Universal Plate, H-Plate; 2-hole through 8-hole plates

Screws: 3.0 and 3.5 locking and non-locking screws

Plate: Available in 1 size only

Screws: 3.5mm locking screw with washer, 4.5mm locking screw with washer or 4.5mm cortical screw Plate Length:3.10in– 4.28in Plate Thickness: .098in -.123in Plate Length: 0.9in – 2.3in Plate Thickness: 0.050-.064in Plate Length: 3.7in Plate Thickness: .138in

The subject CoLink® NeoFuse Plating System is made of Titanium Alloy and has been demonstrated to be substantially equivalent to the previously cleared devices identified above as the products are similar in indications, materials and geometry. Performance Testing: Testing was conducted per a modified ASTM F382 set up for the plates for the CoLink® NeoFuse Plating System and the plates were found to be substantially equivalent. No additional mechanical testing was required for the screws. The CoLink® NeoFuse Plating System were validated per ISO 10993-1 for biocompatibility, ISO 11137-2 for gamma sterilization and ISO 11607-1 for packaging. Endotoxin testing was completed per ANSI/AAMI ST72.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 23, 2026