iSR’obot Mona Lisa 2.0
Biobot Surgical Pte., Ltd.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
iSR’obot Mona Lisa 2.0 is a user -controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third -party needle or a probe, during image -guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Biobot Surgical Pte. Ltd ________________________________________________________________
Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations. The iSR’obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multi -planar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.
Technology Characteristics Compared to Predicate Devices iSR’obot Mona Lisa 2.0 employs the same fundamental scientific technology (design, function and specifications) as that of its legally marketed predicate device, iSR’obot Mona Lisa 1.0 (K203659). Similarities in technology characteristics include:
• Platform-hosted motorized devices and are able to provide 2D and 3D views of the prostate gland;
• Use the same technology to acquire a transrectal ultrasound image to plan and guide a needle for diagnostic procedure;
• Control the trajectory and depth for needle placement via needle guiding mechanism;
• Fusion of Magnetic Resonance and Ultrasound images for mapping planning information of the prostate gland
• Re-positioning of subsequent needle insertion for the same target lesion after the user has indicated the actual needle landing position from the prior insertion.
Modifications to iSR’obot Mona Lisa 2.0 include:
• The iSR’obot Mona Lisa 2.0 is intended for image -guided diagnostic and interventional procedures while the legally marketed predicate device is intended only for image-guided diagnostic procedure;
• Visualization of simulated images in an interventional procedure of 1) the lesions and their margins and 2) treatment zone;
• Needle guide holder, which acts as a channel for insertion tool positioning, is detachable and can be cleaned and sterilized for subsequent use, and enables multi-needle insertion trajectory planning for interventional procedures; Biobot Surgical Pte. Ltd ________________________________________________________________
• Patient movement adjustment function for re-alignment of the prostate model with the live-ultrasound image in the event that prostate shifts due to patient movement during a procedure;
• Connectivity to picture archiving and communication system (PACS) to download DICOM images and download/upload of procedure case; and
• Provide an offline post-procedure review
Other changes include minor user interface variations such as graphics user interface (GUI) design. These differences do not significantly affect the function or use of the device, nor do they raise new or additional safety risks.
These changes are being implemented as a product improvement effort and not due to a corrective action or field action.
iSR’obot Mona Lisa 2.0 has the following technological characteristics, similar to the Artemis (K162474):
• Same intended use for insertion tool planning and navigation for ultrasound image-guided prostate diagnostic and interventional procedures.
• Connecting to standard ultrasound system to display and acquire 2D live ultrasound images of the prostate gland
• Perform other viewing and imaging -processing function such as image segmentation, image multi planar reconstruction, image overlays, a nd measurements
• Image registration and elastic fusion of ultrasound image with previously created MRI prostate model and/or prior biopsy plan on application software
• Generate 2D/3D model of a prostate model and trajectory of insertion tools
• Plan insertion tool positions on the fused prostate model by physicians and mark the actual position of insertion tools
• Connectivity to PACS to download DICOM images and download/upload of procedure case
• Artemis has a needle guide mechanism attachment similar to the iSR ’obot Mona Lisa 2.0, which allows insertion of third-party insertion tools
The differences between iSR’obot Mona Lisa 2.0 and Artemis (K162474) are that:
• Anatomical Access - iSR’obot Mona Lisa 2.0 is intended for insertion tools positioned transperineally while Artemis is intended for insertion tools positioned transperineally and transrectally procedure. Biobot Surgical Pte. Ltd ________________________________________________________________
• Positioning of insertion tools - iSR’obot Mona Lisa 2.0 has an 8 axis motorized robotic arm that navigates the needle guide holder based on the clinician’s insertion tool position plan. The Artemis has a semi-robotic arm (mechanical arm with encoders) that clinicians may control to track and visualize the insertion tool as it is being positioned. Despite the slight difference, the clinician is in control throughout and manually inserts the insertion tool for both devices.
• Adjustment of prostate model in the event of patient movement - iSR’obot Mona Lisa 2.0 provides an adjustment panel for users to manually adjust the prostate model to align with the ultrasound live image of the shifted prostate. Artemis tracks the position and orientation of the ultrasound probe, and users mark landmark features on the ultrasound live image to enable its software to automatically adjust the prostate model to align with the ultrasound live image of the shifted prostate.
• Information for active surveillance - Artemis is intended to be used for patients in active surveillance to keep track of previous procedures information and outcomes. iSR’obot Mona Lisa 2.0 allows previous procedures information to be accessed for patients under active surveillance, but the histopathological outcome is not currently captured in the software.
The technological characteristic above are features that assists physicians to plan and insert an insertion tool. The physician has full control of the tasks to be performed. According to the risk analysis, non -clinical performance tests, device verification and validation tests, the above differences do not result in additional or increase to existing known risks for iSR’obot Mona Lisa 2.0.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。