FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEApril 29, 2022PRODUCT CODEJAK

Indications for use

The BodyTom 64 system is intended to be used for x -ray computed tomography applications for anatomy that can be imaged in the 85cm aperture. The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast. The BodyTom 64 system can be used for low dose lung cancer screening. The screening must be performed in compliance with the approved and established protocols as defined by professional medical societies. *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. Comparison of Technological Characteristics with the Predicate Device The BodyTom 64 , for its intended use, is of comparable type in design, material, functionality, and technology and is substantially equivalent to the cleared predicate device

– BodyTom Elite device (K170238).

We modified the cleared BodyTom Elite (K170238) within our design controls to include technology improvements, which includes hardware updates to the Data Acquisition Board (DAS) to accommodate for the 64 cross sectional CT images produced instead of the previous 32 slices provided by the predicate.

The following differences exist between the subject device ( BodyTom 64 ) and the previously cleared predicate device (BodyTom Elite K170238).

▪ Data Rate: From a rotating component to a stationary component has increased from 1.25 Gb to 4.25 Gb

▪ Slice Thickness: The slice thickness has decreased from 1.25mm to 0. 625mm to provide more detailed images.

▪ Software: o The operating system has been updated to Linux. o The workstation computer software has been updated to account for the new slice thickness values.

▪ Hardware update: The design of the internal hardware of the BodyTom 64 has been updated to accommodate for the 64 image slices offered. o The Data Acq uisition System (DAS) control board has bee n updated to accommodate the new data rate. o The slip ring has been updated for higher contactless data transfer. o Addition of a heat exchanger to the DAS o Updated collimator design to accommodate the 64 image slices.

▪ Accessories: The BodyTom 64 is accompanied by a workstation computer.

The software revision has undergone several updates since the BodyTom Elite (K170238) was cleared. The updates were implemented to enhance the user experience and improve clinical workflow while using the CT system by adding features based on feedback from our current customers , as well as the inclusion of Linux as the operating system . The BodyTom 64 software functions are similar to the predicate BodyTom Elite (K170238) device.

The internal verification and validation activities and external testing of product safety and EMC / EMI was completed successfully. The differences noted above raise no new issues of safety or effectiveness based on all testing performed. Below a summary has been provided for the testing conducted.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 22, 2026